Agili-C CHP0753Recalls & reported problems
Cartiheal (2009) Ltd · 3X Agili-C D7.5 Implant Pack · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290019087175), received between Jan 15, 2026 and Jan 15, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture1
Patient problems
- Arthralgia1
- Swelling/ Edema1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 15, 2026Fracture
IT WAS REPORTED THAT, AFTER A CARTILAGE REPAIR SURGERY HAD BEEN CONDUCTED ON (B)(6) 2025 WITH THREE (3) 7.5 MM X 10 MM AGILI-C IMPLANTS ON AN UNSPECIFIED ANATOMICAL LOCATION, THE PATIENT STARTED EXPERIENCING PAIN AND SWELLING AROUND FOUR (4) MONTHS POST-OPERATIVELY. THE SURGEON NOTICED THAT THE MOST POSTERIOR IMPLANT WAS LEFT PROUD. A REVISION SURGERY WAS CONDUCTED ON (B)(6) 2025 TO REPLACE SUCH IMPLANT WITH A 10 MM X 10 MM AGILI-C IMPLANT. PART OF THE IMPLANT HAD BROKEN OFF AND WAS FLOATING AROUND IN THE JOINT; IT WAS REMOVED WITH AN ARTHROSCOPIC GRASPER. THE SITE OF THE REVISION WAS PREPARED WITH NORMAL AGILI-C TECHNIQUE FOR A 10MM IMPLANT. THE PATIENT HAS PRESENTED POSITIVE IMPROVEMENTS THUS FAR.
Patient problems: Arthralgia; Swelling/ Edema
MAUDE report 24079824 ↗
Other Implant, Resorbable, For Articular Osteochondral Repair devices
- Agili-C CHP0752Cartiheal (2009) Ltd · Implant, Resorbable, For Articular Osteochondral Repair3 reports
- Agili-C CH-125Cartiheal (2009) Ltd · Implant, Resorbable, For Articular Osteochondral Repair2 reports
- Agili-C CHP0755Cartiheal (2009) Ltd · Implant, Resorbable, For Articular Osteochondral Repair1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page