Skip to content

NIRxcell CoCr Coronary Stent on RX System NXL25012USRecalls & reported problems

Medinol Ltd. · Class 3 · Implant

All models of NIRxcell CoCr Coronary Stent on RX System (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290107010764), received between Apr 9, 2017 and Apr 9, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Difficult to Advance1
  • Improper or Incorrect Procedure or Method1
  • Kinked1

Patient problems

  • Aneurysm1
  • Myocardial Infarction1
  • Stenosis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2017Break; Kinked; Improper or Incorrect Procedure or Method; Difficult to Advance; Break; Kinked; Improper or Incorrect Procedure or Method; Difficult to Advance

    REPORTABILITY WAS DECIDED ON MARCH 28, 2017 AFTER RE-REVIEWING THE EVENT FILE. STEMI PATIENT THAT ALSO HAD AN ABDOMINAL AORTIC ANEURYSM WITH MULTIPLE LESIONS IN LAD AND OM1; PHYSICIAN AND STAFF IN THE CASE NOTED EXTREMELY TORTUOUS ANATOMY ESPECIALLY AT THE SITE OF THE UNTREATED LESION, MID OM1; THERE WAS AN EXTREMELY TIGHT TORTUOUS LOOP IN THE OM1 WHERE THE NIRXCELL 2.5 X 12 WAS UNABLE TO TRACK TO DESPITE THE USE OF A GUIDELINER, MULTIPLE PRE DILATIONS, AND DEPLOYMENT OF A 2.5 X 17 IN THE PROXIMAL OM1; PHYSICIAN APPLIED ENOUGH PUSH ON THE SDS OF THE NIRXCELL 2.5 X 12 THAT THE CATHETER KINKED AND EVENTUALLY SNAPPED; THE PATIENT WAS NOT HARMED DURING THE PROCEDURE; ULTIMATELY NO STENTS WERE ABLE TO TRACK TO THIS LESION IN THE MID OM1; THE PHYSICIAN WAS ABLE TO SUCCESSFULLY TREAT A TIGHT LESION IN THE LAD WITH A 3.0 X 17 NIRXCELL AND A DIFFERENT LESION IN THE PROXIMAL OM1 WITH A 2.5 X 17 NIRXCELL. THE DEVICE WAS USED IN ACUTE MI. DE NOVO, STENOSIS 95%, NO BIFURCATION, NO PRE EXISTING CLOT. THERE WAS PREDILATION WITH PRESSURE 12. STENT DILATION PRESSURE WAS 18. THERE WAS POST DILATION WITH PRESSURE 20. TIMI FLOW PRE PROCEDURE WAS 2 AND TIMI FLOW POST PROCEDURE WAS 3. TWO STENTS WERE DE

    Patient problems: Aneurysm; Myocardial Infarction; Stenosis

    MAUDE report 6474624 ↗