Skip to content

NIRxcell CoCr Coronary Stent on RX System NXL35028USRecalls & reported problems

Medinol Ltd. · Class 3 · Implant

All models of NIRxcell CoCr Coronary Stent on RX System (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07290107010986), received between Apr 9, 2017 and Apr 9, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • No Consequences Or Impact To Patient1
  • Stenosis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE STENT CAME OFF THE STENT DELIVERY BALLOON IN THE GUIDE CATHETER AS THE DOCTOR WAS RETRIEVING IT FROM THE OSTIAL RCA BECAUSE IT WOULDN'T CROSS. WHEN HE PULLED THE SDS ALL THE WAY OUT HE NOTICED THE STENT WASN'T ON THE BALLOON. HE THEN USED FLUORO AND DETERMINED THE STENT WAS IN THE GUIDE CATHETER AT WHICH TIME HE HAD TO PULL THE GUIDE AND WIRE OUT OF THE PATIENT. DE NOVO, 85% STENOSIS, NO BIFURCATION, NO PRE EXISTING CLOT. THERE WAS NO PRE DILATION. LESION CALCIFICATION: MEDIUM. VESSEL CALCIFICATION: LOW. VESSEL WAS NOT TORTUOUS. THE PROCEDURE WAS COMPLETED WITH ANOTHER PRODUCT. THERE WAS NO INJURY TO THE PATIENT. ADDITIONAL INFORMATION: THE SYSTEM WAS INSPECTED PRIOR TO USE AND APPEARED NORMAL. THE SYSTEM WAS STORED, HANDLED AND PREPPED ACCORDING TO THE IFU. THE GUIDING CATHETER AND STENT DELIVERY SYSTEM WERE NOT REMOVED AS A SINGLE UNIT. THE DELIVERY SYSTEM WAS REMOVED BY RETRACTING IT INTO THE GUIDING CATHETER.

    Patient problems: No Consequences Or Impact To Patient; Stenosis

    MAUDE report 6474625 ↗

Other Stent, Coronary devices

All Stent, Coronary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page