Skip to content

Paltop 21-70022Recalls & reported problems

Paltop Advanced Dental Solutions Ltd · Dynamic Dental Implant Ø6mm L6mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07290108696363), received between Dec 8, 2025 and May 6, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate3

Patient problems

  • Inadequate Osseointegration3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 6, 2026Failure to Osseointegrate

    FAILED TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25091025 ↗
  • InjuryFeb 3, 2026Failure to Osseointegrate

    FAILED TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24244797 ↗
  • InjuryDec 8, 2025Failure to Osseointegrate; Failure to Osseointegrate

    FAILED TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23738346 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page