Quick summary
This Noris Medical Tuff endosseous root-form dental implant is a titanium screw placed in the jawbone to support a tooth replacement. Most of the 68 reports describe the implant not fusing with the bone, becoming loose or mobile and failing, sometimes with pain, infection, or bone loss around the site. The reports mention 66 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 68
- Deaths reported
- 0
- Injuries reported
- 66
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 07290108812190), received between Oct 5, 2023 and Sep 15, 2026 — classified as 66 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth33
- Failure to Osseointegrate23
- Loss of Osseointegration6
- Insufficient Device Problem Information4
- Adverse Event Without Identified Device or Use Problem1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- Failure of Implant54
- Implant Pain9
- Unspecified Infection7
- Osteolysis6
- Insufficient Information5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.
InjurySep 15, 2026Loosening of Implant Not Related to Bone-Ingrowth
The clinician reports the implant was inserted 2026 (b)(6). On 2026 (b)(6), Loosening of Implant Not Related to bone ingrowth was verified. Observed during the event: Mobility. The device was forwarded to the manufacturer. No patient operative or post-operative complications reported.
Patient problems: Failure of Implant
MAUDE report 50771387 ↗InjurySep 15, 2026Loosening of Implant Not Related to Bone-Ingrowth
The clinician reports the implant was inserted 2026 (b)(6). On 2026 (b)(6), Loosening of Implant Not Related to bone ingrowth was verified. Observed during the event: Mobility. The device was forwarded to the manufacturer. No patient operative or post-operative complications reported.
Patient problems: Failure of Implant
MAUDE report 50771336 ↗InjuryMay 23, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026. ON (B)(6) 2026, THE IMPLANT WAS NOT OSSEOINTEGRATED. LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 25265178 ↗InjuryMay 23, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026. ON (B)(6) 2026, THE IMPLANT WAS NOT OSSEOINTEGRATED. LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 25264792 ↗InjuryApr 5, 2026Insufficient Device Problem Information
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, IMPLANT FAILED WAS REPORTED WITHOUT REASON. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED. BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.
Patient problems: Insufficient Information
MAUDE report 24786656 ↗InjuryMar 31, 2026Insufficient Device Problem Information
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, IMPLANT FAILED WAS REPORTED WITHOUT REASON. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED. BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.
Patient problems: Insufficient Information
MAUDE report 24746669 ↗InjuryMar 24, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 24680949 ↗InjuryMar 24, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 24679423 ↗InjuryMar 17, 2026Loss of Osseointegration
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY. OBSERVED DURING THE EVENT: BONE LOSS/MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Osteolysis
MAUDE report 24618856 ↗InjuryMar 15, 2026Loss of Osseointegration
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: BONE LOSS/MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Osteolysis
MAUDE report 24601616 ↗InjuryJan 29, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Implant Pain; Unspecified Infection
MAUDE report 24208525 ↗InjuryJan 29, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 24208524 ↗InjuryJan 26, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 24177250 ↗InjuryJan 26, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Implant Pain; Unspecified Infection
MAUDE report 24177248 ↗InjuryDec 22, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS ARTEROSCLEROSIS. DOCTOR SAID: IMPLANT FAILED. THE REASON FOR FAILURE IS THAT THE IMPLANTS DID NOT HAVE INITIAL STABILITY. IT WAS OPEN AND NEEDED TO BE REPLACED WITH A WIDER IMPLANT. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 23863397 ↗InjuryDec 16, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS ARTERIOSCLEROSIS. DOCTOR SAID: IMPLANT FAILED. THE REASON FOR FAILURE IS THAT THE IMPLANTS DID NOT HAVE INITIAL STABILITY. IT WAS OPEN AND NEEDED TO BE REPLACED WITH A WIDER IMPLANT. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 23816402 ↗InjuryNov 3, 2025Loss of Osseointegration; Loss of Osseointegration
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: BONE LOSS/MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Implant Pain; Osteolysis; Unspecified Infection
MAUDE report 23450379 ↗InjuryNov 3, 2025Loss of Osseointegration; Loss of Osseointegration
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: BONE LOSS/MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Implant Pain; Osteolysis; Unspecified Infection
MAUDE report 23446854 ↗InjuryAug 7, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY, A SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22737027 ↗InjuryAug 7, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY, A SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22736394 ↗InjuryJun 5, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 22148299 ↗InjuryJun 5, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 22148298 ↗InjuryJun 4, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 22140329 ↗InjuryJun 4, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 22140327 ↗InjuryMay 15, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant
MAUDE report 22026140 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page