Tuff Implant NM-F4216Recalls & reported problems

Noris Medical Ltd · Tuff, D4.2, L16, Int. Hex Ti. Implant · Class 2 · Implant

All models of Tuff Implant (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Noris Medical Tuff dental implant is a root-form screw implant used to anchor replacement teeth in the jaw. The 49 reports mostly describe implants that loosened or never fused with the surrounding bone, with several also mentioning infection around the implant site, pain, or bone loss; 48 reports involved injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
48
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 07290108812206), received between May 5, 2024 and May 25, 2026 — classified as 48 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 11 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 5 reportsApr 2025: 0 reportsMay 2025: 5 reportsJun 2025: 6 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 3 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth24
  • Failure to Osseointegrate18
  • Insufficient Device Problem Information4
  • Loss of Osseointegration2
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • Failure of Implant36
  • Unspecified Infection9
  • Implant Pain4
  • Insufficient Information4
  • Osteolysis2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 25, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25268295 ↗
  • InjuryMay 23, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, THE IMPLANT WAS NOT OSSEOINTEGRATED. LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 25265176 ↗
  • InjuryMay 23, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25264805 ↗
  • InjuryMay 23, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, THE IMPLANT WAS NOT OSSEOINTEGRATED. LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 25264791 ↗
  • InjuryFeb 26, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 24454295 ↗
  • InjuryFeb 22, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 24410525 ↗
  • InjuryFeb 2, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 24230740 ↗
  • InjuryJan 29, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 24208537 ↗
  • InjuryJan 26, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 24177265 ↗
  • InjuryJan 26, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 . ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 24177249 ↗
  • InjuryDec 22, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. DOCTOR SAID: IMPLANT FAILED. THE REASON FOR FAILURE IS THAT THE IMPLANTS DID NOT HAVE INITIAL STABILITY. IT WAS OPEN AND NEEDED TO BE REPLACED WITH A WIDER IMPLANT. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23863414 ↗
  • InjuryDec 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 . ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 23837536 ↗
  • InjuryDec 17, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. DOCTOR SAID: IMPLANT FAILED. THE REASON FOR FAILURE IS THAT THE IMPLANTS DID NOT HAVE INITIAL STABILITY. IT WAS OPEN AND NEEDED TO BE REPLACED WITH A WIDER IMPLANT. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23827733 ↗
  • InjuryDec 15, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 23802868 ↗
  • InjuryJun 11, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/DIABETES. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22183594 ↗
  • InjuryJun 11, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22183574 ↗
  • InjuryJun 10, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/DIABETES. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22175266 ↗
  • InjuryJun 10, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22175256 ↗
  • InjuryJun 5, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22148297 ↗
  • InjuryJun 4, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 . ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22140330 ↗
  • InjuryMay 14, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4216 LOT#9000425 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22017551 ↗
  • MalfunctionMay 14, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4216 LOT#9001124 IMPLANT WAS OPENED IN THE PACKAGE DUE TO TRANSPORTATION. THE IMPLANT WASN'T USED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22017549 ↗
  • InjuryMay 14, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4216 LOT#9000425 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 22017455 ↗
  • InjuryMay 7, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4216 LOT#9001124 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #4. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION, DIABETES, CARDIAC PROBLEMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21975994 ↗
  • InjuryMay 7, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4216 LOT # 9001124 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #4. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION, DIABETES, CARDIAC PROBLEMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21975942 ↗