Skip to content

Tuff Implant NM-F4220Recalls & reported problems

Noris Medical Ltd · Tuff, D4.2, L20, Int. Hex Ti. Implant · Class 2 · Implant

All models of Tuff Implant (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07290108812220), received between Sep 25, 2024 and Apr 5, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information4
  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Failure to Osseointegrate1

Patient problems

  • Insufficient Information4
  • Failure of Implant3
  • Implant Pain1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 5, 2026Insufficient Device Problem Information

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, IMPLANT FAILED WAS REPORTED WITHOUT REASON. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED. BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.

    Patient problems: Insufficient Information

    MAUDE report 24786658 ↗
  • InjuryApr 5, 2026Insufficient Device Problem Information

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, IMPLANT FAILED WAS REPORTED WITHOUT REASON. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED. BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.

    Patient problems: Insufficient Information

    MAUDE report 24786657 ↗
  • InjuryMar 31, 2026Insufficient Device Problem Information

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, IMPLANT FAILED WAS REPORTED WITHOUT REASON. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED. BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.

    Patient problems: Insufficient Information

    MAUDE report 24746670 ↗
  • InjuryMar 31, 2026Insufficient Device Problem Information

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, IMPLANT FAILED WAS REPORTED WITHOUT REASON. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED. BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.

    Patient problems: Insufficient Information

    MAUDE report 24746668 ↗
  • InjuryFeb 26, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 24454294 ↗
  • InjuryFeb 22, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 24410529 ↗
  • InjuryJan 19, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2024, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Implant Pain

    MAUDE report 21187726 ↗
  • InjurySep 25, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4220 LOT#9000193 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #19 DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS OSTEOPOROSIS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 20298993 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page