Skip to content

Tuff TT Implant NM-F4311Recalls & reported problems

Noris Medical Ltd · Tuff TT, D4.2, L11.5, Int. Hex Ti. Implant · Class 2 · Implant

All models of Tuff TT Implant (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Noris Medical Tuff TT implant is a root-form dental implant. Reports mostly describe implants loosening or failing to bond with the bone, with pain, implant failure and infection mentioned. There are 24 reports, with no deaths and 24 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
22
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 07290108812268), received between Jun 25, 2024 and Mar 2, 2026 — classified as 22 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate10
  • Loosening of Implant Not Related to Bone-Ingrowth9
  • Loss of Osseointegration3

Patient problems

  • Failure of Implant21
  • Unspecified Infection7
  • Implant Pain6
  • Osteolysis3

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 2, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS BRUXISM. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain; Unspecified Infection

    MAUDE report 24481900 ↗
  • InjuryFeb 24, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS BRUXISM. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain; Unspecified Infection

    MAUDE report 24428145 ↗
  • InjuryNov 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 . ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23582183 ↗
  • InjuryNov 17, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23567017 ↗
  • InjurySep 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION/PREDIABETIC. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY.THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 23083130 ↗
  • InjuryJun 11, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: INFECTION/MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain; Unspecified Infection

    MAUDE report 22183593 ↗
  • InjuryJun 10, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS DIABETES. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: INFECTION/MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain; Unspecified Infection

    MAUDE report 22175267 ↗
  • InjuryMay 14, 2025Loss of Osseointegration; Loss of Osseointegration

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: BONE LOSS, MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 22017555 ↗
  • InjuryMay 14, 2025Loss of Osseointegration; Loss of Osseointegration

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: BONE LOSS, MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 22017418 ↗
  • InjuryMar 31, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000793 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3 DUE TO BONE TOO SOFT. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION, THYROID THERAPY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21729429 ↗
  • InjuryMar 31, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000793 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3 DUE TO BONE TOO SOFT. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION, THYROID THERAPY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21729373 ↗
  • InjuryFeb 23, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000793 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #27, DUE TO BONE TOO SOFT. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21442059 ↗
  • InjuryFeb 23, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000793 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #27, DUE TO BONE TOO SOFT. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21442052 ↗
  • InjuryFeb 20, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT# 9000793 IMPLANT LACKED PRIMARY STABILITY DUE TO LOW TORQUE, SOFT BONE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, THE PATIENT HAS DIABETES. AT THE EVENT, THE PATIENT EXPERIENCED: INFECTION, BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 21419955 ↗
  • InjuryFeb 20, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT# 9000793 IMPLANT LACKED PRIMARY STABILITY DUE TO LOW TORQUE, SOFT BONE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, THE PATIENT HAS DIABETES. AT THE EVENT, THE PATIENT EXPERIENCED: INFECTION, BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 21419825 ↗
  • InjuryOct 13, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2024, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: INFECTION, MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 20435712 ↗
  • InjurySep 25, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT BONE TOO SOFT AND NM-F4311 LOT#9000746 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #28 . THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS THYROID PROBLEMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 20300378 ↗
  • InjurySep 25, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000793 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #30 DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS OSTEOPOROSIS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 20298995 ↗
  • InjuryJul 21, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000589 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 19788552 ↗
  • InjuryJul 7, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4311 LOT#9000330 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 19686202 ↗
  • InjuryJun 25, 2024Loss of Osseointegration; Loss of Osseointegration

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER, THE PATIENT HAS PERIODONTITIS. OBSERVED DURING THE EVENT: BONE LOSS, MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 19604254 ↗
  • InjuryJun 25, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2024, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Implant Pain

    MAUDE report 19604252 ↗

Other Implant, Endosseous, Root-Form devices