Skip to content

Zygomatic Implant NM-F4455Recalls & reported problems

Noris Medical Ltd · Zygomatic, D4.2, L55, Int. Hex Ti. Implant · Class 2 · Implant

All models of Zygomatic Implant (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07290108812398), received between Jun 6, 2024 and Mar 5, 2026 — classified as 14 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate6
  • Loosening of Implant Not Related to Bone-Ingrowth4
  • Fracture2
  • Insufficient Device Problem Information2

Patient problems

  • Failure of Implant11
  • Implant Pain3
  • Unspecified Infection2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 5, 2026Fracture

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, IMPLANT FRACTURE WAS VERIFIED. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: IMPLANT FRACTURE/MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 24520548 ↗
  • InjuryMar 5, 2026Fracture

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, IMPLANT FRACTURE WAS VERIFIED. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: IMPLANT FRACTURE/ MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 24520502 ↗
  • InjurySep 14, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, PREVIOUS DRUG USER. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23047739 ↗
  • InjurySep 11, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER, PREVIOUS DRUG USER. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23025188 ↗
  • InjuryApr 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4455 LOT#0025757 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21917619 ↗
  • InjuryApr 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4455 LOT #0025757 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21917533 ↗
  • InjuryJan 16, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4455 LOT#0026705 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21166027 ↗
  • InjurySep 1, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4455 LOT#0025757 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #15 DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 20115343 ↗
  • InjuryJul 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4455 LOT#0026705 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #3. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 19803708 ↗
  • InjuryJul 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2024, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 19803702 ↗
  • InjuryJul 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023. ON (B)(6) 2024, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: INFECTION, MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 19803653 ↗
  • InjuryJul 21, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F4455 LOT #: 0023952 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 19788554 ↗
  • InjuryJun 16, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    A HEALTHCARE PROFESSIONAL REPORTED THE WAY THEY PLANNED IT THE PLATFORM WAS 3-4MM SUPRACRESTAL WE HAD TO SWITCH TO SHORTER IMPLANT FOR RESTORABILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19544726 ↗
  • InjuryJun 6, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE HEALTHCARE PROFESSIONAL REPORTS THAT THE NM-F4455 LOT#0025757 IMPLANT WAS INSERTED IN THE PATIENT'S MOUTH IN (B)(6) 2024. THE IMPLANT WAS REMOVED DURING PLACEMENT DUE TO PTERYGOID SOFT BONE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. A LONGER IMPLANT WAS REPLACED. NO PATIENT-OPERATIVE OR POST-OPERATIVE COMPLICATIONS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19474693 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page