Tuff TT Implant NM-F5113Recalls & reported problems
Noris Medical Ltd · Tuff TT, D5.0, L13, Int. Hex Ti. Implant · Class 2 · Implant
All models of Tuff TT Implant (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 18
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 07290108812510), received between May 12, 2024 and Jun 2, 2026 — classified as 18 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate8
- Loosening of Implant Not Related to Bone-Ingrowth6
- Insufficient Device Problem Information3
- Loss of Osseointegration1
Patient problems
- Failure of Implant12
- Unspecified Infection4
- Insufficient Information3
- Implant Pain2
- Osteolysis1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
InjuryJun 2, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED.ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Implant Pain
MAUDE report 25357586 ↗InjuryJun 1, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED.ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Implant Pain
MAUDE report 25349380 ↗InjurySep 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 23082082 ↗InjurySep 17, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 23072085 ↗InjuryJul 6, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22408480 ↗InjuryJul 3, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A HEALTHY/ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22394287 ↗InjuryMay 6, 2025Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT SURGEON UNABLE TO GET APPROPRIATE TORQUE ON IMPLANT SITE #4, NM-F5113, LOT#9000592 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 21967745 ↗InjuryMay 6, 2025Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT SURGEON UNABLE TO GET APPROPRIATE TORQUE ON IMPLANT SITE #4, NM-F5113, LOT#9000592 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 21967645 ↗InjuryApr 1, 2025Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F5113 LOT#9000592 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #6 DUE TO BONE TOO SOFT. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, THE PATIENT HAS THYROID THERAPY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 21738544 ↗InjuryApr 1, 2025Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F5113 LOT#9000592 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #6 DUE TO BONE TOO SOFT. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, THE PATIENT HAS THYROID THERAPY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 21738429 ↗InjuryJul 8, 2024Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F5113 LOT#9000592 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 19688569 ↗InjuryJul 1, 2024Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F5113 LOT#0024835 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #29 DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 19641191 ↗InjuryJun 26, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
A HEALTH PROFESSIONAL REPORTS THE IMPLANT FAILED, HOWEVER BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.
Patient problems: Insufficient Information
MAUDE report 19613921 ↗InjuryJun 26, 2024Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F5113 LOT#9000592 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 19613860 ↗InjuryJun 25, 2024Loss of Osseointegration; Loss of Osseointegration
THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: Osteolysis
MAUDE report 19604237 ↗InjuryMay 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
A HEALTH PROFESSIONAL REPORTS THE IMPLANT FAILED, HOWEVER BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED
Patient problems: Insufficient Information
MAUDE report 19319516 ↗InjuryMay 15, 2024Failure to Osseointegrate; Failure to Osseointegrate
A HEALTHCARE PROFESSIONAL REPORTED THAT NM-F5113 LOT#9000592 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANTATION AND REMOVAL DATE HAS NOT BEEN PROVIDED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Patient problems: Failure of Implant
MAUDE report 19319515 ↗InjuryMay 12, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
A HEALTH PROFESSIONAL REPORTS THE IMPLANT FAILED, HOWEVER BASED ON THE AVAILABLE INFORMATION THE EXACT ISSUE COULD NOT BE IDENTIFIED.
Patient problems: Insufficient Information
MAUDE report 19295543 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page