This Noris Medical PteryFit dental implant is a root-form screw implant placed in the jawbone. The 50 adverse event reports mostly describe implants that fail to fuse with the bone and later loosen, become mobile, or break, with patients reporting pain, infection and bone loss around the implant. There were 50 injuries and no deaths, and no recalls name this device. Reports fell from 18 in the earlier 12 months to 5 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
50
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 07290108817133), received between May 5, 2024 and Jun 2, 2026 — classified as 50 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 11 reportsJun 2024: 5 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 6 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 5 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate27
  • Loosening of Implant Not Related to Bone-Ingrowth16
  • Insufficient Device Problem Information3
  • Loss of Osseointegration3
  • Fracture1

Patient problems

  • Failure of Implant42
  • Implant Pain7
  • Osteolysis3
  • Unspecified Infection3
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 2, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS A HYPERTENSION/CANCER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 25357597 ↗
  • InjuryJun 1, 2026Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2026. ON (B)(6) 2026, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS A HYPERTENSION/CANCER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 25349374 ↗
  • InjuryMar 8, 2026Fracture

    THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT DAY ON (B)(6) 2025. OBSERVED DURING THE EVENT: IMPLANT BROKE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THE MALFUNCTION DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 24541738 ↗
  • InjuryNov 4, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 23464977 ↗
  • InjuryNov 3, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 23449941 ↗
  • InjuryOct 15, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 . ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23300430 ↗
  • InjuryOct 15, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 23298796 ↗
  • InjuryAug 10, 2025Loss of Osseointegration; Loss of Osseointegration

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT HAS HIGH CHOLESTEROL/OSTEOPENOSIS. OBSERVED DURING THE EVENT: BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Osteolysis

    MAUDE report 22767829 ↗
  • InjuryAug 10, 2025Loss of Osseointegration; Loss of Osseointegration

    THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT HAS HIGH CHOLESTEROL/OSTEOPENOSIS. OBSERVED DURING THE EVENT: BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Osteolysis

    MAUDE report 22767812 ↗
  • InjuryMay 12, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED ON TOOTH #13. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 22001872 ↗
  • InjuryMay 12, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED ON TOOTH #4. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 22001868 ↗
  • InjuryMay 12, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED ON TOOTH #4. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 22001704 ↗
  • InjuryMay 12, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED ON TOOTH #13. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 22001701 ↗
  • InjuryMar 24, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED ON TOOTH #15. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, THE PATIENT HAS HEP C. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21676253 ↗
  • InjuryMar 24, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 . ON (B)(6) 2025 , LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED ON TOOTH #15. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER, THE PATIENT HAS HEP C. OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21676198 ↗
  • InjuryFeb 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9000949 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21410253 ↗
  • InjuryFeb 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9001158 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21410209 ↗
  • InjuryFeb 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9001158 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21410039 ↗
  • InjuryFeb 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9000949 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21409968 ↗
  • InjuryFeb 13, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9000949 IMPLANT LACKED PRIMARY STABILITY DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21373527 ↗
  • InjuryJan 26, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT # 9000949 IMPLANT LACKED PRIMARY STABILITY DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21236988 ↗
  • InjuryJan 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9000649 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #16 DUE TO LOW TORQUE. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21187718 ↗
  • InjuryJan 16, 2025Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9001043 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 21166032 ↗
  • InjuryOct 6, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT#9000949 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 20379902 ↗
  • InjuryOct 6, 2024Failure to Osseointegrate; Failure to Osseointegrate

    A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4222 LOT# 0023358 IMPLANT LACKED PRIMARY STABILITY. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.

    Patient problems: Failure of Implant