Slimline 550 FiberRecalls & reported problems
Lumenis Ltd. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07290109140025), received between Apr 13, 2021 and Aug 24, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Failure to Deliver Energy1
- Overheating of Device1
- Smoking1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2022Break; Smoking; Use of Device Problem; Break; Smoking; Use of Device Problem
IT WAS REPORTED THAT, DURING PREPARATION FOR A CERVICAL CONIZATION PROCEDURE, THIS FIBER WAS LEFT ON A CLOTH THAT HAD BEEN PLACED ON THE PATIENT. THE CONSOLE SCREEN STATUS REMAINED IN READY MODE WHILE THE LASER FIBER WAS OUTSIDE THE PATIENT. THE CLOTH SUDDENLY CAUGHT FIRE; THE FIRE WAS QUICKLY EXTINGUISHED AND THERE HAS NO HARM TO THE PATIENT OR MEDICAL STAFF. THE PROCEDURE WAS COMPLETED WITH THE SAME FIBER. IT IS BELIEVED THAT THE SYSTEM WAS MISTAKENLY ACTIVATED, LEADING TO THE FIRE. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE HANDPIECE SHOWED BURN MARKS, INDICATING THAT THE LASER HAD BEEN FIRED INSIDE THE HANDPIECE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15285739 ↗MalfunctionOct 11, 2021Failure to Deliver Energy; Failure to Deliver Energy
LUMENIS WAS INFORMED BY A FOREIGN DISTRIBUTOR THAT A FOREIGN USER FACILITY REPORTED THAT DURING A HOLEP PROCEDURE IN WHICH A LUMENIS PULSE 120H LASER SYSTEM WITH A LUMENIS SLIMLINE 550¿ WAS BEING UTILIZED, "45 MINUTES AFTER THE PROCEDURE BEGAN, THE MACHINE MADE WEIRD NOISES WHILE LASING". THE USER ELECTED TO STOP AND THE REMAINDER OF THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY A TURP PROCEDURE. NO REPORT OF PATIENT COMPLICATIONS WAS RECEIVED, AND NO REPORT WAS RECEIVED ALLEGING THE DEVICE MALFUNCTION CAUSED OR CONTRIBUTED TO ANY CHANGE IN THE PATIENT'S CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12603813 ↗MalfunctionApr 13, 2021Overheating of Device; Overheating of Device
A FOREIGN USER FACILITY REPORTED THAT DURING A PULMONARY PROCEDURE IN WHICH A LUMENIS VERSAPULSE POWERSUITE 80/100 LASER SYSTEM WITH A LUMENIS SLIMLINE 550¿ WAS BEING UTILIZED, SPARKS/FLAME BACKFIRED LEADING TO IRREPARABLE SCOPE DAMAGE. THE FIBER AND SCOPE WERE REMOVED FROM THE PATIENT AND AFTER A 15 MINUTES DELAY THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE. NO REPORT OF PATIENT COMPLICATIONS WAS RECEIVED, AND NO REPORT WAS RECEIVED ALLEGING THE DEVICE MALFUNCTION CAUSED OR CONTRIBUTED TO ANY CHANGE IN THE PATIENT'S CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11654905 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.