Skip to content

Slimline EZ SIS 365 FiberRecalls & reported problems

Lumenis Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07290109140209), received between Feb 4, 2020 and Jan 19, 2024 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Defective Device1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Burn(s)1
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 19, 2024Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DISTAL END OF THE LASER FIBER BROKE OFF. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE PATIENT AND PROCEDURE OUTCOME TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18544980 ↗
  • InjuryJan 9, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT DURING A PROCEDURE, THERE WAS A DEVICE MALFUNCTION AND DUE TO THIS, THE PHYSICIAN GOT INJURED. THE PATIENT AND PROCEDURE OUTCOME ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18476374 ↗
  • MalfunctionSep 17, 2023Break; Break

    IT WAS REPORTED THAT DURING AN URETEROSCOPY PROCEDURE FOR THE TREATMENT OF STONES, THE FIBER BROKE WHILE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A SECOND FIBER. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17762342 ↗
  • MalfunctionApr 17, 2023Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT URETER AND BLADDER STONES, THE FIBER PHYSICALLY BROKE. IT WAS NOTED THAT NO SIGNIFICANT BEND DEFLECTION WAS REQUIRED TO REACH THE TREATMENT SITE. A SECOND FIBER WAS USED; HOWEVER, THIS FIBER TIP OVERHEATED AND BURNT AND NEEDED TO BE REPLACED BY A THIRD FIBER TO COMPLETE THE PROCEDURE. THERE WERE NO OPERATOR OR PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16756685 ↗
  • MalfunctionMar 17, 2023Break; Break

    IT WAS REPORTED THAT THE LASER FIBER SNAPPED AND BROKE OUTSIDE THE PATIENT DURING THE PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16567810 ↗
  • InjuryFeb 4, 2020Fracture; Fracture

    A FOREIGN USER FACILITY REPORTED THAT AT THE END OF A URETERIC STONE FRAGMENTATION PROCEDURE IN WHICH A LUMENIS PULSE 100 LASER WAS BEING UTILIZED WITH A LUMENIS SLIMLINE EZ 365 FIBER, THE FIBER FRACTURED CAUSING A BURN TO A MEMBER OF THE NURSING STAFF. THE STAFF BURN WAS REPORTED TO HAVE "OCCURRED AT THE END OF THE PROCEDURE; WHEREAS THE LASER WAS NO LONGER NEEDED AFTER THAT SO THE PROCEDURE WAS NOT AFFECTED."

    Patient problems: Burn(s)

    MAUDE report 9664099 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page