Skip to content

Slimline EZ SIS 550 FiberRecalls & reported problems

Lumenis Ltd. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07290109140216), received between Apr 26, 2023 and Jun 13, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Defective Device1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Burn(s)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 13, 2025Break; Break

    IT WAS REPORTED THAT DURING A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) FOR BENIGN PROSTATIC HYPERPLASIA, THE FIBER BROKE AT INSERTION POINT TO SCOPE. A SECOND FIBER WAS TRIED, AND IT BROKE TOO. DUE TO THIS, THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS. THIS REPORT IS FOR THE SECOND BROKEN FIBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206882 ↗
  • MalfunctionJun 13, 2025Break; Break

    IT WAS REPORTED THAT DURING A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) FOR BENIGN PROSTATIC HYPERPLASIA, THE FIBER BROKE AT INSERTION POINT TO SCOPE. A SECOND FIBER WAS TRIED, AND IT BROKE TOO. DUE TO THIS, THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS. THIS COMPLAINT IS FOR THE FIRST BROKEN FIBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206610 ↗
  • MalfunctionJun 7, 2024Break; Break

    IT WAS REPORTED THAT DURING THE PROCEDURE THE FIBER WAS PLACED INSIDE THE OPENED ENDED URETERAL CATHETER AND BEING ACTIVELY USED WHEN THE LASER FIBER ITSELF BROKE IN TWO PIECES OUTSIDE THE PATIENT. A LOUD NOISE WAS AUDIBLE IN THE ROOM. THE LASER WAS TURNED OFF AND THE TWO PIECES OF THE LASER FIBER WERE RETRIEVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT COMPLICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19489460 ↗
  • InjuryApr 26, 2023Defective Device; Insufficient Device Problem Information; Defective Device; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING A PROCEDURE, THE ENERGY IN THE FIBER FIRED INTO THE HAND OF THE PHYSICIAN, CAUSING A BURN; NO INFORMATION WAS PROVIDED ON WHETHER OR NOT TREATMENT WAS ADMINISTERED FOR THE BURN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT PATIENT COMPLICATIONS. FURTHER INFORMATION OBTAINED INDICATED THAT THE USER IS EXPERIENCING ISSUES WITH THE SCOPES (NON- BSC) USED DURING PROCEDURE. THE SHORT-TERM SOLUTION WITH THE SCOPE ISSUES WAS TO INSERT A SLIM CATHETER INTO THE WORKING CHANNEL OF THE SCOPE TO PREVENT THE LASER RATTLING AROUND DURING PROCEDURE. HOWEVER, THE WORKING CHANNEL APPEARS TO BE TOO LARGE IN DIAMETER.

    Patient problems: Burn(s)

    MAUDE report 16820568 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page