Xpeeda DSL FiberRecalls & reported problems
Lumenis Ltd. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07290109140278), received between Feb 20, 2020 and Feb 18, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Material Fragmentation1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 18, 2025Break; Break
IT WAS REPORTED THAT DURING A CYSTOLITHOLAPAXY PROCEDURE, THE STAFF NOTICED THAT THE FIBER WIRE BROKE IN HALF. WHEN THE FIBER BROKE HIT THE PATIENT DRAPE AND BURNED THE DRAPE. THERE WAS A BARRIER BETWEEN THE PATIENT AND THE DRAPE, THEREFORE, THERE WAS NO PATIENT HARM. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE METHOD WITHOUT PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21406369 ↗MalfunctionJun 12, 2024Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FIBER TIP BROKE OFF WHEN INTRODUCED TO THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19519458 ↗MalfunctionNov 9, 2023Break; Break
IT WAS REPORTED THAT THE LASER FIBER CAP BROKE DURING THE HOLMIUM LASER ENUCLEATION OF THE PROSTATE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18101196 ↗MalfunctionJun 28, 2023Break; Break
IT WAS REPORTED THAT THE FIBER DETERIORATED DURING THE PROCEDURE AND AT THE END OF THE PROCEDURE, THE TIP DETACHED. THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17221913 ↗MalfunctionNov 17, 2022Break; Break
IT WAS REPORTED THAT DURING A PROSTATE ABLATION PROCEDURE, THE FIBER TIP BROKE OFF INSIDE THE PATIENT BLADDER. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER WITHOUT PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15813313 ↗MalfunctionOct 19, 2022Break; Thermal Decomposition of Device; Detachment of Device or Device Component; Break; Thermal Decomposition of Device; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PROCEDURE THE LASER TIP FELL OFF AND EXPLODED INSIDE THE PATIENT. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15633123 ↗MalfunctionFeb 20, 2020Material Fragmentation; Material Fragmentation
A USER FACILITY REPORTED THAT DURING A PROCEDURE IN WHICH A LUMENIS XPEEDA FIBER WAS BEING UTILIZED , THE TIP OF THE FIBER BROKE OFF IN THE PATIENT. THE TIP WAS RETRIEVED AND A NEW FIBER WAS USED TO COMPLETE THE CASE. NO REPORT OF PATIENT INJURY WAS RECEIVED, AND THE EVENT DID NOT CAUSE OR CONTRIBUTE TO ANY CHANGE IN THE PATIENT'S CONDITION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9731660 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page