Skip to content

VersaPulse PowerSuite 80WRecalls & reported problems

Lumenis Ltd. · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07290109140599), received between Sep 20, 2022 and Nov 14, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Output below Specifications” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output below Specifications2
  • Device Displays Incorrect Message1
  • No Display/Image1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2024Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT DURING A PROCEDURE, THERE WAS A WEAK OUTPUT OBSERVED, THERE WAS NOT ANY ERROR DISPLAYED. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE PATIENT OUTCOME TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20685296 ↗
  • MalfunctionOct 4, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING AN INSPECTION OF THE DEVICE UPON ITS RETURN, THE EMERGENCY BUTTON WAS NOT WORKING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20370540 ↗
  • MalfunctionJan 2, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT, THE CONSOLE SCREEN WAS BLANK, AND NOTHING WAS DISPLAYED WHEN TRIED TO USE THE DEVICE. IT WAS REBOOTED SEVERAL TIMES WITH NO IMPROVEMENT, SO THE CASE WAS DISCONTINUED. THE PATIENT WAS STABLE UNDER GENERAL ANESTHESIA. THERE WAS NO PATIENT COMPLICATION. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18426080 ↗
  • MalfunctionMar 3, 2023Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT THE VERSAPULSE DID NOT START DURING A PROCEDURE. THE ERROR 221 WAS SHOWED IN THE SCREEN. THE PATIENT WAS UNDER GENERAL ANESTHESIA AND THE PROCEDURE WAS CANCELLED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16479782 ↗
  • MalfunctionSep 20, 2022Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT DURING THE PROCEDURE THE CONSOLE BECAME UNSTABLE. AFTER REBOOTING, THE SCREEN TURNED RED. THE PROCEDURE WAS CANCELLED AND NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT SEDATION STATUS OF UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15457583 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.