Skip to content

OtoLase™ FiberRecalls & reported problems

Lumenis Ltd. · Class 2

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07290109141862), received between Aug 23, 2023 and Jun 12, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Adverse Event Without Identified Device or Use Problem2
  • Defective Device2
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE FIBER WAS RETURN TO BOSTON SCIENTIFIC WITHOUT PERFORMANCE ALLEGATION DESCRIPTION. UPON COMPLETION OF THE FIBER ANALYSIS, IT WAS IDENTIFIED THAT FIBER TIP WAS BROKEN. THEREFORE, BASED ON THIS SPECIFIC ANALYSIS FINDING, REPORTABILITY CRITERIA IS MET. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22191455 ↗
  • MalfunctionMay 15, 2025Break; Break

    IT WAS REPORTED THAT, DURING A STAPEDECTOMY SURGERY, THE FIBER BURNT OUT AND BROKE IN THE MIDDLE CREATING FIRE. LASER WAS IN DEFAULT SETTING WITH POWER 10 W. THE CONSOLE SEEMED TO BE WORKING FINE. THE PHYSICIAN IMMEDIATELY CHANGED THE LASER. LATER THE PROCEDURE WAS COMPLETED USING A NON-FIBER MODE (FREE HAND) WITH A MICROMANIPULATOR. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22031891 ↗
  • MalfunctionMay 7, 2025Break; Break

    IT WAS REPORTED THAT THE FIBER BURNT AND BROKE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21981385 ↗
  • MalfunctionMar 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A FIBER WAS RETURNED WITH NO ALLEGATION. ANALYSIS OF THE RETURN DEVICE IDENTIFIED THAT THE FIBER TIP WAS BROKEN. THERE WAS NO FURTHER INFORMATION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21606393 ↗
  • MalfunctionOct 10, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT DURING PROCEDURE, THE POWER WAS NOT HIGH ENOUGH TO COMPLETE THE PROCEDURE. IT WAS FOUND THAT THE ISSUE WAS WITH THE FIBER. FIBER TIPS WERE REPLACED BUT THE ISSUE CONTINUED. DIFFERENT CONNECTIONS AND SETTING WERE ATTEMPTED BUT IT DID NOT WORK. THERE WERE NO OTHER FIBERS AVAILABLE, AND THE PROCEDURE WAS CANCELLED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20423283 ↗
  • MalfunctionJun 27, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING PREPARATION OF A PROCEDURE THE FIBER WAS FIRED TWICE, HOWEVER, LIGHT WAS NOT EMITTED FROM THE END OF THE FIBER. THE FIBER WAS TRIED ON TWO DIFFERENT LASER CONSOLES BUT WAS NOT TRIED ON THE PATIENT. THE FIBER WAS REPLACED TO COMPLETE THE PROCEDURE. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED A FIBER CONNECTOR FRACTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629643 ↗
  • MalfunctionMay 8, 2024Break; Break

    IT WAS REPORTED THAT, DURING REMOVAL ATTEMPTS OF THREE FIBERS, THEY GOT STUCK AND SUBSEQUENTLY BROKE INSIDE THE HANDPIECES, DAMAGING BOTH THE FIBERS AND THE HANDPIECES AND RENDERING THEM UNUSABLE. A THOROUGH INSPECTION OF THE LASER EQUIPMENT AND ALIGNMENT OF THE DUO SYSTEM SHOWS NO ISSUES. THERE WAS NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR FIBER 2 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272424 ↗
  • MalfunctionMay 8, 2024Break; Break

    IT WAS REPORTED THAT, DURING REMOVAL ATTEMPTS OF THREE FIBERS, THEY GOT STUCK AND SUBSEQUENTLY BROKE INSIDE THE HANDPIECES, DAMAGING BOTH THE FIBERS AND THE HANDPIECES AND RENDERING THEM UNUSABLE. A THOROUGH INSPECTION OF THE LASER EQUIPMENT AND ALIGNMENT OF THE DUO SYSTEM SHOWS NO ISSUES. THERE WAS NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR FIBER 3 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272421 ↗
  • MalfunctionMay 8, 2024Break; Break

    IT WAS REPORTED THAT, DURING REMOVAL ATTEMPTS OF THREE FIBERS, THEY GOT STUCK AND SUBSEQUENTLY BROKE INSIDE THE HANDPIECES, DAMAGING BOTH THE FIBERS AND THE HANDPIECES AND RENDERING THEM UNUSABLE. A THOROUGH INSPECTION OF THE LASER EQUIPMENT AND ALIGNMENT OF THE DUO SYSTEM SHOWS NO ISSUES. THERE WAS NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR FIBER 1 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19272414 ↗
  • MalfunctionAug 23, 2023Defective Device; Defective Device

    IT WAS REPORTED THAT THE FIBER WAS STUCK ON THE DEVICE AND THE AIMING BEAM WAS NOT PRESENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER FIBER WITHOUT PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED A FRACTURED CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17608577 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →