FiberLase™ ENDURE CO2 FiberRecalls & reported problems
Lumenis Ltd. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07290109142081), received between Apr 21, 2023 and Dec 15, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Component Missing” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Component Missing2
- Defective Device2
- Firing Problem2
- Use of Device Problem2
- Melted1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2025Melted; Nonstandard Device; Melted; Nonstandard Device
IT WAS REPORTED THAT AT THE END OF A PROCEDURE, THE INSULATION FIBER PART, APPROXIMATELY 30CM FROM THE PROXIMAL PART OF THE FIBER, HAD MELTED. THE PROCEDURE WAS COMPLETED WITH ORIGINAL FIBER WITHOUT PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23802436 ↗MalfunctionApr 21, 2023Use of Device Problem; Component Missing; Defective Device; Firing Problem; Use of Device Problem; Component Missing; Defective Device; Firing Problem
IT WAS REPORTED THAT AFTER THE FIRST USE OF THE FIBER, SOMETHING AFFECTED THE DEVICE DURING THE CLEANING AND STERILIZATION PROCESS WITHIN THE FACILITY. IT WAS NOTED THAT THE DEVICE HAD BEEN LEFT IN THE DRYER ALL NIGHT AT AROUND 50 C OF CONTINUOUSLY APPLIED HEAT, ALTHOUGH THE RECOMMENDED DRYING TIME ON MANUAL IS 20 MINUTES. AN ON-SITE CHECK WAS PERFORMED AND IDENTIFIED CORRUGATED MARKS THAT APPEARED TO BE SCRATCHES ON THE FIBER SURFACE. IT WAS NOTED THAT THE MANUAL FOR THE DEVICES WAS NOT RECEIVED. ADDITIONALLY, IT WAS INDICATED THAT THE AIMING BEAM REACHED THE TIP, BUT THE LASER WOULD NOT FIRE. THE EVENT OCCURRED BEFORE THE PROCEDURE; HOWEVER, THE ANESTHESIA HAD ALREADY BEEN ADMINISTERED. THE PROCEDURE WAS NOT COMPLETED DUE TO THE SAME DEVICE BEING UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED DUE TO PROCEDURE ABORTED AFTER SEDATION WAS ADMINISTERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16788323 ↗MalfunctionApr 21, 2023Use of Device Problem; Component Missing; Defective Device; Firing Problem; Use of Device Problem; Component Missing; Defective Device; Firing Problem
IT WAS REPORTED THAT AFTER THE FIRST USE OF THE FIBER, SOMETHING AFFECTED THE DEVICE DURING THE CLEANING AND STERILIZATION PROCESS WITHIN THE FACILITY. IT WAS NOTED THAT THE DEVICE HAD BEEN LEFT IN THE DRYER ALL NIGHT AT AROUND 50 C OF CONTINUOUSLY APPLIED HEAT, ALTHOUGH THE RECOMMENDED DRYING TIME ON MANUAL IS 20 MINUTES. AN ON-SITE CHECK WAS PERFORMED AND IDENTIFIED CORRUGATED MARKS THAT APPEARED TO BE SCRATCHES ON THE FIBER SURFACE. IT WAS NOTED THAT THE MANUAL FOR THE DEVICES WAS NOT RECEIVED. ADDITIONALLY, IT WAS INDICATED THAT THE AIMING BEAM REACHED THE TIP, BUT THE LASER WOULD NOT FIRE. THE EVENT OCCURRED BEFORE THE PROCEDURE; HOWEVER, THE ANESTHESIA HAD ALREADY BEEN ADMINISTERED. THE PROCEDURE WAS NOT COMPLETED DUE TO THE SAME DEVICE BEING UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED DUE TO PROCEDURE ABORTED AFTER SEDATION WAS ADMINISTERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16788317 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026