This Lumenis Stellar M22 is a laser and light-based device used for skin treatments such as facial and pigmentation procedures. It has 23 adverse event reports, all involving injuries — mostly burns to patients during treatment, with some reports describing a strong flash, unusual noise, or an error code before the burn; no deaths have been reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
22
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 07290109144726), received between Feb 14, 2021 and Aug 9, 2026 — classified as 22 injuries.

The most frequently reported specific device problem is “Defective Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Defective Component4
  • Use of Device Problem4
  • Use of Incorrect Control/Treatment Settings4
  • Insufficient Device Problem Information3
  • No Apparent Adverse Event1

Patient problems

  • Burn(s)20
  • Insufficient Information2
  • Skin Erosion1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 9, 2026Use of Incorrect Control/Treatment Settings

    Lumenis received an adverse event report on a patient who sustained injury following treatment by the Stellar device.

    Patient problems: Burn(s)

    MAUDE report 50188829 ↗
  • InjuryMay 26, 2026Insufficient Device Problem Information

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED INJURY FOLLOWING TREATMENT BY STELLAR DEVICE. PER THE CUSTOMER, PATIENT HAD MELASMA TREATMENT AND SHE ENDED UP GETTING MORE PIGMENT (WORSE).

    Patient problems: Insufficient Information

    MAUDE report 25277614 ↗
  • InjuryApr 19, 2026Defective Component

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY STELLAR DEVICE. ACCORDING TO THE CUSTOMER, THE ISSUE OCCURRED WHILE USING THE STELLAR M22, FACIAL A, 560 NM, AT 10 J. ON THE 17TH SHOT, AN UNUSUAL NOISE CAME FROM THE MAIN UNIT, FOLLOWED BY A STRONG FLASH. AFTER THE IRRADIATION, A BURN WAS OBSERVED. DUE TO THE BURN SYMPTOMS, THE PATIENT WAS EXAMINED AND OINTMENT WAS PROVIDED.

    Patient problems: Burn(s)

    MAUDE report 24921127 ↗
  • InjuryApr 19, 2026Defective Component

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY STELLAR DEVICE. ACCORDING TO THE CUSTOMER, THE ISSUE OCCURRED WHILE USING THE STELLAR M22, FACIAL A, 560 NM, AT 10 J. ON THE 17TH SHOT, AN UNUSUAL NOISE CAME FROM THE MAIN UNIT, FOLLOWED BY A STRONG FLASH. AFTER THE IRRADIATION, A BURN WAS OBSERVED. DUE TO THE BURN SYMPTOMS, THE PATIENT WAS EXAMINED AND OINTMENT WAS PROVIDED.

    Patient problems: Burn(s)

    MAUDE report 24921126 ↗
  • InjuryMar 22, 2026Use of Incorrect Control/Treatment Settings

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED INJURY FOLLOWING TREATMENT BY THE STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 24658127 ↗
  • InjuryMar 15, 2026Defective Component

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY STELLAR DEVICE. PER THE CUSTOMER, THE FIRST PULSE DURING TREATMENT CREATED A STRONG FLASH AND IT BURNED THE PATIENT THEN THE ERROR CODE 1082 (SYSTEM FPGA FAULT) APPEARED ON THE SCREEN.

    Patient problems: Burn(s)

    MAUDE report 24601650 ↗
  • InjuryMar 15, 2026Defective Component

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY STELLAR DEVICE. PER THE CUSTOMER, THE FIRST PULSE DURING TREATMENT CREATED A STRONG FLASH AND IT BURNED THE PATIENT THEN THE ERROR CODE 1082 (SYSTEM FPGA FAULT) APPEARED ON THE SCREEN.

    Patient problems: Burn(s)

    MAUDE report 24601649 ↗
  • InjuryNov 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON SEVERAL PATIENTS WHO SUSTAINED SWELLING FOLLOWING TREATMENT BY STELLAR DEVICE. HOWEVER, SINCE THE CUSTOMER DIDN'T PROVIDE US EVEN ONE INCIDENT FORM, THIS CASE WILL BE EVALUATED IN CURRENT SINGLE COMPLAINT.

    Patient problems: Insufficient Information

    MAUDE report 23507765 ↗
  • InjuryAug 25, 2025Use of Device Problem; Use of Device Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PERSON WHO SUSTAINED INJURY FOLLOWING TREATMENT BY STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 22882273 ↗
  • InjuryAug 25, 2025Use of Device Problem; Use of Device Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PERSON WHO SUSTAINED INJURY FOLLOWING TREATMENT BY STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 22882258 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED INJURY ON THE LEFT LATERAL FOREHEAD USING IPL HANDPIECE FOLLOWING TREATMENT BY STELLAR DEVICE. THIS IS A SECOND INJURY OF THE SAME PATIENT (USER FACILITY HERSELF - DR. (B)(6)) THAT HAS BEEN HANDLED THROUGH COMPLAINT #(B)(4). THIS WAS DURING ANOTHER TREATMENT USING DIFFERENT SETTINGS ON A DIFFERENT DAY.

    Patient problems: Burn(s)

    MAUDE report 22369226 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED INJURY ON THE LEFT LATERAL FOREHEAD USING IPL HANDPIECE FOLLOWING TREATMENT BY STELLAR DEVICE. THIS IS A SECOND INJURY OF THE SAME PATIENT (USER FACILITY HERSELF - DR. (B)(6) THAT HAS BEEN HANDLED THROUGH COMPLAINT # (B)(4). THIS WAS DURING ANOTHER TREATMENT USING DIFFERENT SETTINGS ON A DIFFERENT DAY.

    Patient problems: Burn(s)

    MAUDE report 22369160 ↗
  • InjuryJun 15, 2025Use of Device Problem; Use of Device Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED INJURY USING IPL HANDPIECE FOLLOWING TREATMENT BY STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 22208583 ↗
  • InjuryJun 15, 2025Use of Device Problem; Use of Device Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED INJURY USING IPL HANDPIECE FOLLOWING TREATMENT BY STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 22208580 ↗
  • InjuryJul 18, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN USING IPL HANDPIECE FOLLOWING TREATMENT BY STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 19770786 ↗
  • InjuryFeb 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON SEVERAL PATIENTS WHO SUSTAINED BURNS FOLLOWING TREATMENT BY STELLAR DEVICE. SINCE LUMENIS DIDN'T RECEIVE AN INCIDENT FORM, THIS CASE HANDLED AS ONE COMPLAINT.

    Patient problems: Burn(s)

    MAUDE report 18737572 ↗
  • InjurySep 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LUMENIS RECEIVED ON (B)(6), 2023 AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED A BURN ON HER LEG WITH OPEN WOUND FOLLOWING TREATMENT BY STELLAR DEVICE. ACCORDING TO THE INFORMATION RECEIVED, THE EVENT OCCURRED ON (B)(6), 2022. LUMENIS CANNOT CONFIRM EVER RECEIVING NOTIFICATION OF THIS EVENT AT THAT TIME. SEE "COMPLAINT DESCRIPTION" ABOVE FOR THEIR FULL NOTIFICATION.

    Patient problems: Burn(s); Skin Erosion; Skin Inflammation/ Irritation

    MAUDE report 17716894 ↗
  • InjuryApr 4, 2023No Apparent Adverse Event; No Apparent Adverse Event

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON SEVERAL PATIENTS WHO SUSTAINED BURNS FOLLOWING TREATMENT BY STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 16674189 ↗
  • InjuryMar 23, 2023Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY STELLAR M22 DEVICE.

    Patient problems: Burn(s)

    MAUDE report 16596825 ↗
  • InjuryMar 23, 2023Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY STELLAR M22 DEVICE.

    Patient problems: Burn(s)

    MAUDE report 16596773 ↗
  • InjuryAug 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURN FOLLOWING TREATMENT BY THE M22 STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 15293087 ↗
  • InjuryFeb 14, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    LUMENIS RECEIVED AN ADVERSE EVENT REPORT ON A PATIENT WHO SUSTAINED BURNS ON AN UNKNOWN AREA FOLLOWING TREATMENT WITH A LUMENIS STELLAR DEVICE.

    Patient problems: Burn(s)

    MAUDE report 11325998 ↗

FDA clearance and classification

The FDA regulates product code ONG (powered laser surgical instrument with microbeam\fractional output) under 21 CFR 878.4810 as a Class II device. 21 CFR 878.4810 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Stellar M22

Has the Stellar M22 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07290109144726), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Stellar M22?

The FDA's MAUDE database lists 22 adverse event reports for this device, received between Feb 14, 2021 and Aug 9, 2026. In those reports, the most common specific device problems are “Defective Component” (4), “Use of Device Problem” (4) and “Use of Incorrect Control/Treatment Settings” (4); the most common patient problems are “Burn(s)” (20), “Skin Erosion” (1) and “Skin Inflammation/ Irritation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Stellar M22?

None of the reports are classified as deaths, 22 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Stellar M22, and is it still sold?

The FDA's device database lists Lumenis Ltd. as the company responsible for it (brand Stellar M22, model Stellar M22). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Powered Laser Surgical Instrument With Microbeam\Fractional Output devices

All Powered Laser Surgical Instrument With Microbeam\Fractional Output devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page