Skip to content

Lumenis Pulse 120H (REF GA-2003477)Recalls & reported problems

Lumenis Ltd. · Class 2

Adverse event reports
14
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07290109145440), received between Feb 24, 2023 and Sep 17, 2026 — classified as 1 injury, 13 malfunctions.

The most frequently reported specific device problem is “Output below Specifications” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output below Specifications9
  • Nonstandard Device5
  • Device Displays Incorrect Message3
  • Adverse Event Without Identified Device or Use Problem2
  • Defective Device1
  • Noise, Audible1
  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Nonstandard Device; Output below Specifications

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812423 ↗
  • MalfunctionAug 18, 2026Output below Specifications

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50304666 ↗
  • MalfunctionMay 15, 2026Nonstandard Device; Output below Specifications

    IT WAS REPORTED THAT DURING PROCEDURE, THE CONSOLE SEEMED TO HAVE LOW OUTPUT. THERE IS NO INFORMATION ABOUT PATIENT AND PROCEDURE OUTCOMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25191459 ↗
  • MalfunctionMar 2, 2026Nonstandard Device; Output below Specifications

    IT WAS REPORTED THAT DURING PROCEDURE THE OUTPUT WAS WEAK. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24491392 ↗
  • MalfunctionJan 13, 2026Nonstandard Device

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE DEBRIS SHIELD BURNED OUT TWICE, LEADING TO THE CANCELLATION OF THE PROCEDURE. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24060640 ↗
  • MalfunctionSep 12, 2025Poor Quality Image; Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Noise, Audible; Poor Quality Image; Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Noise, Audible

    IT WAS REPORTED DURING PROCEDURE, WHEN LASER IRRADIATION WAS ATTEMPTED, A NOISE WAS HEARD AFTER THE DOTTED LINE WAS DISPLAYED. THE DEBRIS SHIELD AND FIBER WERE REPLACED, HOWEVER, THERE WAS NO IMPROVEMENT. THE PROCEDURE WAS CANCELLED. THERE WERE NO PATIENT COMPLICATIONS REPORTED. PATIENT WAS UNDER GENERAL ANESTHESIA. IT WAS CONFIRMED THAT THE ISSUE WAS PRIMARILY DUE TO THE CONSOLE. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23039060 ↗
  • InjuryMay 1, 2025Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT, DURING LITHOTRIPSY PROCEDURE, THE CONSOLE DISPLAYED ERRORS 188: "COOLING FLOW SWITCH ERROR" AND 55: "COMMUNICATION PROTOCOL ERROR". THE LASER COULD NOT BE USED, AND A URETERAL STENT WAS PLACED. THE PROCEDURE HAD TO BE CANCELLED/RESCHEDULED. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21947308 ↗
  • MalfunctionOct 30, 2024Adverse Event Without Identified Device or Use Problem; Output below Specifications; Adverse Event Without Identified Device or Use Problem; Output below Specifications

    IT WAS REPORTED THAT DURING A PROCEDURE, THE CONSOLE WAS IN LOW OUTPUT MODE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20570810 ↗
  • MalfunctionNov 16, 2023Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT THE CONSOLE DISPLAYED ERROR CODES 48 (SIS-FIBER IS NOT LUMENIS FIBER (OR NO SIS DETECTED)), 58 (SELF-TEST PROCESS FAILURE (ONE OF THE TESTS COMPLETED WITH FAIL STATUS)), 150 (LASER DECK - FIBER NOT CONNECTED) AND 188(COOLING SYSTEM ERROR-COOLING FLOW SWITCH ERROR). THE CONSOLE WAS REBOOTED BUT IT DID NOT RESOLVE THE ISSUE AND THE PROCEDURE WAS CANCELLED. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE PATIENT OUTCOME TO DATE, DESPITE GOOD FAITH EFFORTS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18155011 ↗
  • MalfunctionNov 16, 2023Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT THE CONSOLE DISPLAYED ERROR CODES 48 (SIS-FIBER IS NOT LUMENIS FIBER (OR NO SIS DETECTED)), 58 (SELF-TEST PROCESS FAILURE (ONE OF THE TESTS COMPLETED WITH FAIL STATUS)), 150 (LASER DECK - FIBER NOT CONNECTED) AND 188 (COOLING SYSTEM ERROR-COOLING FLOW SWITCH ERROR). THE CONSOLE WAS REBOOTED BUT IT DID NOT RESOLVE THE ISSUE, AND THE PROCEDURE WAS CANCELLED. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE PATIENT OUTCOME TO DATE, DESPITE GOOD FAITH EFFORTS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18154900 ↗
  • MalfunctionNov 8, 2023Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT THE CONSOLE OUTPUT IS LOW. THERE WAS NO PATIENT OR PROCEDURE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18097746 ↗
  • MalfunctionOct 10, 2023Defective Device; Output below Specifications; Defective Device; Output below Specifications

    IT WAS REPORTED THAT THE BLAST SHIELD LENS WAS BURNT DURING THE PROCEDURE AND THE BLAST SHIELD LENS AND PROBE WERE REPLACED BUT THE OUTPUT SEEMS WEAK. THE PROCEDURE WAS COMPLETED WITH A NON-BSC PRODUCT WITHOUT PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17909371 ↗
  • MalfunctionSep 25, 2023Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT DURING A HOLMIUM LASER ENUCLEATION OF THE PROSTATE PROCEDURE TO ENLARGED PROSTATE, THE CONSOLE PRESENTED LOW POWER MODE. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE, WITHOUT PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17808129 ↗
  • MalfunctionFeb 24, 2023Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT DURING PREPARATION FOR THE PROCEDURE, BUT AFTER THE PATIENT HAD BEEN SEDATED, THE CONSOLE WENT INTO LOW POWER MODE. THE PROCEDURE WAS CANCELLED DUE TO THE DEVICE ISSUES. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16435139 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.