Lumenis Pulse 30H (REF GA-2003322)Recalls & reported problems
Lumenis Ltd. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07290109145983), received between Jan 20, 2023 and Aug 11, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Output below Specifications” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Output below Specifications3
- Defective Device2
- Device Displays Incorrect Message2
- Adverse Event Without Identified Device or Use Problem1
- Communication or Transmission Problem1
- Incomplete or Inadequate Connection1
- Noise, Audible1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Output below Specifications
It was reported that during a Capital Equipment Service, it was found a decrease in output value due to age-related degradation of the Brick. There was no patient involved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50202729 ↗MalfunctionMar 30, 2026Incomplete or Inadequate Connection
IT WAS REPORTED THAT, DURING PROCEDURE AND AFTER SEDATION OF THE PATIENT, THE CONSOLE ISSUED ERROR CODE 4 (FOOTSWITCH FAULT). THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24714965 ↗MalfunctionMar 16, 2026Output below Specifications
IT WAS REPORTED THAT DURING PREPARATIONS, THE CONSOLE SEEMS TO HAVE LOW POWER. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24603889 ↗MalfunctionDec 23, 2025Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT DURING PREPARATION POST-SEDATION, THE CONSOLE DISPLAYED ERROR CODES 49 PYRO MAIN NOT EQUAL TO PYRO SAFE AND ERROR CODE 54 SW AUTOCAL FAULT: N=1 / NOT DONE: N=2. THE PROCEDURE WAS DISCONTINUED DUE TO DEVICE DEFECT. THERE WAS NO PATIENT COMPLICATION. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23880421 ↗MalfunctionDec 10, 2025Output below Specifications; Output below Specifications
IT WAS REPORTED THAT THE DEVICE'S LASER OUTPUT WAS DECREASING. THERE WAS NO PATIENT OR PROCEDURE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23774355 ↗MalfunctionOct 29, 2025Smoking; Noise, Audible; Smoking; Noise, Audible
IT WAS REPORTED THAT A LOUD NOISE WAS HEARD FROM THE DEVICE, FOLLOWED BY SMOKE FROM THE REAR PART. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23420548 ↗MalfunctionAug 13, 2024Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ERROR 17: "OVERHEATING ALERT! OVER TEMPERATURE (49 C) THERMAL SWITCH" DISPLAYED. THE FAN SEEMED TO NOT BE ROTATING. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT WHOSE SEDATION STATUS WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19978276 ↗MalfunctionAug 9, 2024Overheating of Device; Defective Device; Overheating of Device; Defective Device
IT WAS REPORTED THAT DURING A PROCEDURE, THE FAN WAS NOT SPINNING, AND AN OVERHEATING ERROR WAS DISPLAYED. DUE TO THIS, THE PROCEDURE WAS NOT COMPLETED. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT HAS BEEN REPORTED WITH A PROCEDURE WHICH ANESTHESIA OR SEDATION STATUS IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19950408 ↗MalfunctionJun 30, 2023Defective Device; Defective Device
IT WAS REPORTED THAT CONSOLE WAS UNABLE TO BE USED AS THE READY BUTTON COULD NOT BE PRESSED EVEN IF THE FIBER IS CONNECTED. THE CONSOLE WAS REBOOTED BUT STILL DID NOT WORK, LEADING TO THE CANCELLATION OF THE PROCEDURE. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT WHOSE SEDATION STATUS WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17240205 ↗MalfunctionJan 20, 2023Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT ERROR CODE 2- SHUTTER ERROR, OCCURRED. THE PROCEDURE WAS ABORTED AND RESCHEDULED FOR THE FOLLOWING WEEK. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16215014 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page