Skip to content

VersaPulse Powersuite VersaPulse Powersuite 80/100W (REF 0638-807-01)Recalls & reported problems

Lumenis Ltd. · Class 2

All models of VersaPulse PowerSuite (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07290109146225), received between Dec 13, 2022 and Feb 19, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Displays Incorrect Message5
  • Output below Specifications3
  • Smoking2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2026Nonstandard Device; Smoking

    IT WAS REPORTED THAT THE DEBRIS SHIELD IS BURNT AND SMOKE WAS RISING. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24385436 ↗
  • MalfunctionMay 30, 2025Smoking; Defective Device; Smoking; Defective Device

    IT WAS REPORTED THAT, THE BLAST SHIELD WAS BURNT AND BURNING SMELL WAS NOTICED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22118485 ↗
  • MalfunctionFeb 21, 2025Adverse Event Without Identified Device or Use Problem; Output below Specifications; Adverse Event Without Identified Device or Use Problem; Output below Specifications

    IT WAS REPORTED THAT THE LASER OUTPUT WAS WEAK. THE CONSOLE DID NOT DISPLAY A COURTESY MESSAGE NOR ERROR CODE. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21437100 ↗
  • MalfunctionAug 23, 2024Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING A CASE, THE CONSOLE DISPLAYED ERROR 505 (VOLTAGE ON HVPS CAPACITOR BANK IS OUTSIDE SPECIFICATION.), IT COULD NOT BE USED EVEN AFTER RESTARTING. DUE TO THIS, ANOTHER DEVICE WAS USED BUT THE STONE WAS IMPACTED IN THE URETER AND THE APPROACH WAS IMPOSSIBLE, SO THE PROCEDURE WAS WITHDRAWN. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR AN ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20054922 ↗
  • MalfunctionMay 22, 2024Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT THE CONSOLE PRESENTED LOW OUTPUT MODE DURING ITS USE. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE PATIENT AND PROCEDURE OUTCOME TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19372259 ↗
  • MalfunctionDec 14, 2023Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE SCREEN DISPLAYED ATTACH FOOTSWITCH WHEN THE FOOTSWITCH WAS STEPPED ON. THE PROCEDURE WAS DISCONTINUED AND THERE WAS NO PATIENT COMPLICATION. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18326194 ↗
  • MalfunctionNov 7, 2023Device Displays Incorrect Message; Output below Specifications; Device Displays Incorrect Message; Output below Specifications

    IT WAS REPORTED THAT, AFTER THE PATIENT HAD BEEN SEDATED FOR A PROCEDURE, THE CONSOLE DISPLAYED A RED SCREEN, INDICATING CASE SAVER MODE. AN ERROR CODE "ALL LASER MALFUNCTION" WAS ALSO DISPLAYED. THE PROCEDURE WAS CANCELLED. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR AN ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18083440 ↗
  • MalfunctionOct 12, 2023Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING PREPARATION OF THE CONSOLE FOR A TRANSURETHRAL URETEROLITHOTRIPSY PROCEDURE ERROR CODE 201 (SHUTTER TEST FAILED) DISPLAYED ON THE CONSOLE SCREEN. THE CONSOLE WAS REBOOTED, BUT THE ISSUE PERSISTED. THERE WERE NO PATIENT COMPLICATIONS, BUT THE PROCEDURE WAS CANCELLED/POSTPONED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17919437 ↗
  • MalfunctionDec 13, 2022Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT ERROR F205: "NEITHER OPTOCOUPLERS FOR THE SHUTTER SEES LIGHT." DISPLAYED. THE PROCEDURE COULD NOT BE COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT WHOSE SEDATION STATUS WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15971828 ↗