Skip to content

CGuard Prime Carotid Stent System CND1040Recalls & reported problems

Inspire M.D Ltd · CGuard Prime Carotid Stent System 135cm,10mmx40mm · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07290120281783), received between Nov 4, 2025 and Aug 17, 2026 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (P240029); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 1, 2026Class IIOpen, Classified

    Stent delivery system may encounter complications during deployment which may result in high resistance or inability to deploy the stent.

    Z-2330-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Activation Failure2
  • Difficult or Delayed Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 17, 2026Difficult or Delayed Activation

    It was reported that during an elective transfemoral carotid artery stenting procedure, the operator applied "extreme pressure" of th CGuard Prime stent delivery system to deploy the stent. The stent was able to be deployed successfully, however the stent was deployed covering a portion of the lesion. A secondary InspireMD stent was used to cover the lesion and the procedure was completed without any patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50276333 ↗
  • InjuryAug 5, 2026Activation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50148448 ↗
  • InjuryAug 5, 2026Activation Failure

    It was reported during a trans ulnar access case, the physician experienced a non-deployment using the CGuard Prime Carotid Stent System 135cm. The physician attempted to use a secondary CGuard Prime Carotid Stent System 135cm using the trans femoral approach, however the issue persisted. A competitor stent was used to complete the procedure successfully with no reported patient harm. The system was able to be deployed outside of the patients body.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50148447 ↗
  • MalfunctionDec 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THE FOLLOWING: PHYSICIAN WAS ABLE TO COMPLETE THE PROCEDURE. PATIENT IS DOING WELL; HOWEVER, THE NOSE OF THE STENT POST DEPLOYMENT WA GETTING STUCK PRIOR RO POST DILATION. HE USED A 3.5 BALLOON FREE IN A MODERATELY CALCIFIED LESION, FOLLOWED WITH A 5.5 BALLOON. DURING THE DEPLOYMENT PROCESS THE STENT DID JUMP BUT DID NOT MIS THE (STENOSIS) LESION, HOWEVER IT DID NOT DEPLOY AT THE DESIRED LOCATION AND HAD TO EXPAND THE STENT. THE PHYSICIAN USED A COMPETITOR STENT TO COVER THE PROXIMAL COMMON CAROTID. THE PHYSICIAN DID NOT EXPERIENCE RESISTANCE. THE NOSE CONE IS BULKY PER PHYSICIAN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23805266 ↗
  • MalfunctionDec 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THE FOLLOWING: PHYSICIAN WAS ABLE TO COMPLETE THE PROCEDURE. PATIENT IS DOING WELL; HOWEVER, THE NOSE OF THE STENT POST DEPLOYMENT WA GETTING STUCK PRIOR RO POST DILATION. HE USED A 3.5 BALLOON FREE IN A MODERATELY CALCIFIED LESION, FOLLOWED WITH A 5.5 BALLOON. DURING THE DEPLOYMENT PROCESS THE STENT DID JUMP BUT DID NOT MIS THE (STENOSIS) LESION; HOWEVER, IT DID NOT DEPLOY AT THE DESIRED LOCATION AND HAD TO EXPAND THE STENT. THE PHYSICIAN USED A COMPETITOR STENT TO COVER THE PROXIMAL COMMON CAROTID. THE PHYSICIAN DID NOT EXPERIENCE RESISTANCE. THE NOSE CONE IS BULKY PER PHYSICIAN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23803692 ↗
  • MalfunctionNov 4, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN INTERVENTION OF A PATIENT EXPERIENCING CLOTS, DISSECTION AND CALCIFICATION IN THE CAROTID, THE PHYSICIAN DECIDED TO USE A CND1040. THE PHYSICIAN WAS ABLE TO ADVANCE INTO THE LESION, BUT DURING DEPLOYMENT, RESISTANCE WAS EXPERIENCED, AND THE STENT DID NOT DEPLOY. THE PHYSICIAN DECIDED NOT TO CONTINUE APPLYING FORCE INTO THE DELIVERY SYSTEM AND PROCEEDED TO REMOVE THE CATHETER/STENT. THE PHYSICIAN FINISHED THE PROCEDURE WITH A COMPETITOR STENT. NO HARM TO THE PATIENT REPORTED. AFTER THE PROCEDURE, THE PHYSICIAN TRIED TO DEPLOY THE STENT IN A FLAT TABLE. STENT DEPLOYED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23466394 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P240029. One submission often covers many models, so check the product and lot details in each record.

  • Z-2329-2026 · May 1, 2026 · Open, Classified
    CGuard Prime Carotid Stent System, 135cm, 10mmx30mm Model/Catalog Number: CND1030
  • Z-2327-2026 · May 1, 2026 · Open, Classified
    CGuard Prime Carotid Stent System, 135cm, 9mx30mm Model/Catalog Number: CND0930
  • Z-2325-2026 · May 1, 2026 · Open, Classified
    Brand Name: CGuard Prime Carotid Stent System, 135cm, 8mmx30mm Model/Catalog Number: CND0830
  • Z-2326-2026 · May 1, 2026 · Open, Classified
    Brand Name: CGuard Prime Carotid Stent System, 135cm, 8mmx40mm Model/Catalog Number: CND0840
  • Z-2328-2026 · May 1, 2026 · Open, Classified
    Brand Name: CGuard Prime Carotid Stent System, 135cm, 9mx40mm Model/Catalog Number: CND0940

Other Stent, Carotid devices

All Stent, Carotid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page