Ventilator Trolley 007384Recalls & reported problems
Breas Medical AB · Ventilator Trolley (base unit) · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07321820073849), received between Dec 9, 2021 and Dec 9, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 21, 2022Class IITerminated
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
Z-0943-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
Patient problems
- Ecchymosis1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 9, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
NOTE: VOLUNTARY REPORT OF NON-SERIOUS ADVERSE EVENT. ON (B)(6) 2021, BREAS WAS NOTIFIED THAT A VIVO 50 VENTILATOR FELL OFF A TROLLEY STAND ONTO PATIENT'S FOOT, CAUSING A BRUISE WHICH DID NOT REQUIRE MEDICAL INTERVENTION. THE TROLLEY STAND FOR BREAS VENTILATOR WAS MISSING A METAL WASHER WHICH RENDERED THE MOUNTING BRACKET FOR VENTILATOR UNSTABLE.
Patient problems: Ecchymosis
MAUDE report 12967503 ↗
FDA clearance and classification
The FDA regulates product code FOX (stand, infusion) under 21 CFR 880.6990 as a Class I device. 21 CFR 880.6990 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Ventilator Trolley 007384
Has the Ventilator Trolley 007384 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0943-2022 (Class II), began on Mar 21, 2022: Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Ventilator Trolley 007384?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 9, 2021. That report lists device problem “Detachment of Device or Device Component” (1); patient problem “Ecchymosis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Ventilator Trolley 007384?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Ventilator Trolley 007384 FDA approved?
Devices of this type (product code FOX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Breas Medical AB.
Who makes the Ventilator Trolley 007384, and is it still sold?
The FDA's device database lists Breas Medical AB as the company responsible for it (brand Ventilator Trolley, model 007384). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stand, Infusion devices
- NOV - Novum IQ 40700HUBBaxter Healthcare Corporation · STAND, INFUSION1 report
- MEDLINE MDS80600Medline Industries, Inc. · STAND, INFUSION1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page