Compressor Mini 115V 60Hz 6481779Recalls & reported problems
Maquet Critical Care AB · Class 2
Quick summary
This Maquet Compressor Mini, a portable air compressor, has 108 adverse event reports, with no deaths and no injuries. Many reports describe the compressor losing output or pressure, failing to power up, or having moisture problems, and some describe a drainage valve that did not work or leaks. Reports fell to 3 in the last 12 months from 21 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 108
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 07325710001653), received between Jan 5, 2023 and May 12, 2026 — classified as 108 malfunctions.
The most frequently reported specific device problem is “Output Problem” (37 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem37
- Failure to Power Up24
- Moisture or Humidity Problem19
- Gas Output Problem18
- Decrease in Pressure13
- Contamination3
- Gas/Air Leak2
- Increase in Pressure2
Patient problems
- No Clinical Signs, Symptoms or Conditions108
Recent adverse event reports
The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 12, 2026Moisture or Humidity Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25151247 ↗MalfunctionJan 16, 2026Decrease in Pressure; Gas/Air Leak
IT WAS REPORTED THAT THE COMPRESSOR HAD A LEAKAGE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24093885 ↗MalfunctionDec 19, 2025Output Problem; Output Problem
IT WAS REPORTED THAT THE COMPRESSOR MALFUNCTIONED DUE TO EXCESSIVE DUST ACCUMULATION CAUSED BY THE DUST FILTER. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23847827 ↗MalfunctionOct 16, 2025Output Problem; Output Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23308635 ↗MalfunctionAug 22, 2025Gas Output Problem; Gas Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22871607 ↗MalfunctionAug 20, 2025Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem
DURING THE INSPECTION OF THE COMPRESSOR IT WAS DISCOVERED THAT THE DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22848053 ↗MalfunctionJul 17, 2025Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22529236 ↗MalfunctionJul 15, 2025Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22509878 ↗MalfunctionJul 15, 2025Gas Output Problem; Gas Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22506306 ↗MalfunctionJul 1, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22370059 ↗MalfunctionJun 18, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem
DURING THE INSPECTION OF THE COMPRESSOR IT WAS DISCOVERED THAT DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22274157 ↗MalfunctionJun 2, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22123176 ↗MalfunctionMay 28, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
MANUFACTURER'S REF #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22095066 ↗MalfunctionMay 9, 2025Gas Output Problem; Gas Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21992518 ↗MalfunctionMay 7, 2025Gas Output Problem; Gas Output Problem
IT WAS REPORTED THAT THE COMPRESSOR HAD AN ISSUE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21976308 ↗MalfunctionApr 1, 2025Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21737961 ↗MalfunctionMar 21, 2025Output Problem; Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21662155 ↗MalfunctionFeb 21, 2025Gas Output Problem; Gas Output Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21433248 ↗MalfunctionFeb 21, 2025Gas Output Problem; Gas Output Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21432592 ↗MalfunctionFeb 18, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21395975 ↗MalfunctionJan 7, 2025Gas Output Problem; Gas Output Problem
IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21091253 ↗MalfunctionDec 27, 2024Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem
IT WAS REPORTED THAT THE COMPRESSOR'S AIR OUTLET WAS CONTAMINATED WITH WATER. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21027498 ↗MalfunctionDec 13, 2024Failure to Power Up; Moisture or Humidity Problem; Failure to Power Up; Moisture or Humidity Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20923582 ↗MalfunctionNov 7, 2024Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20628925 ↗MalfunctionOct 4, 2024Gas Output Problem; Gas Output Problem
MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20369389 ↗
Other Compressor, Air, Portable devices
- Medline HCS70004Medline Industries, Inc. · Compressor, air, portable386 reports
- Medline HCS70004RDMedline Industries, Inc. · Compressor, air, portable108 reports
- Life2000 MS-01-0093Welch Allyn, Inc. · Compressor, air, portable1 recall1 death5 reports
- Life2000 MS-01-0121Welch Allyn, Inc. · Compressor, air, portable1 recall0 reports
- Life2000 MS-01-0125Welch Allyn, Inc. · Compressor, air, portable1 recall0 reports
- InnoSpire Elegance 00Respironics Respiratory Drug Delivery (Uk) Ltd · Compressor, air, portable19 reports
- InnoSpire Essence 00Respironics Respiratory Drug Delivery (Uk) Ltd · Compressor, air, portable18 reports
- Medline HCSPUPPYMedline Industries, Inc. · Compressor, air, portable15 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.