Skip to content

Compressor Mini 115V 60Hz 6481779Recalls & reported problems

Maquet Critical Care AB · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Maquet Compressor Mini, a portable air compressor, has 108 adverse event reports, with no deaths and no injuries. Many reports describe the compressor losing output or pressure, failing to power up, or having moisture problems, and some describe a drainage valve that did not work or leaks. Reports fell to 3 in the last 12 months from 21 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
108
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 07325710001653), received between Jan 5, 2023 and May 12, 2026 — classified as 108 malfunctions.

The most frequently reported specific device problem is “Output Problem” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 19 reportsJan 2024: 5 reportsFeb 2024: 4 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 4 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem37
  • Failure to Power Up24
  • Moisture or Humidity Problem19
  • Gas Output Problem18
  • Decrease in Pressure13
  • Contamination3
  • Gas/Air Leak2
  • Increase in Pressure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions108

Recent adverse event reports

The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 12, 2026Moisture or Humidity Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25151247 ↗
  • MalfunctionJan 16, 2026Decrease in Pressure; Gas/Air Leak

    IT WAS REPORTED THAT THE COMPRESSOR HAD A LEAKAGE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093885 ↗
  • MalfunctionDec 19, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR MALFUNCTIONED DUE TO EXCESSIVE DUST ACCUMULATION CAUSED BY THE DUST FILTER. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23847827 ↗
  • MalfunctionOct 16, 2025Output Problem; Output Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23308635 ↗
  • MalfunctionAug 22, 2025Gas Output Problem; Gas Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22871607 ↗
  • MalfunctionAug 20, 2025Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem

    DURING THE INSPECTION OF THE COMPRESSOR IT WAS DISCOVERED THAT THE DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22848053 ↗
  • MalfunctionJul 17, 2025Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22529236 ↗
  • MalfunctionJul 15, 2025Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22509878 ↗
  • MalfunctionJul 15, 2025Gas Output Problem; Gas Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22506306 ↗
  • MalfunctionJul 1, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22370059 ↗
  • MalfunctionJun 18, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem

    DURING THE INSPECTION OF THE COMPRESSOR IT WAS DISCOVERED THAT DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22274157 ↗
  • MalfunctionJun 2, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22123176 ↗
  • MalfunctionMay 28, 2025Moisture or Humidity Problem; Moisture or Humidity Problem

    MANUFACTURER'S REF #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095066 ↗
  • MalfunctionMay 9, 2025Gas Output Problem; Gas Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21992518 ↗
  • MalfunctionMay 7, 2025Gas Output Problem; Gas Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR HAD AN ISSUE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21976308 ↗
  • MalfunctionApr 1, 2025Moisture or Humidity Problem; Moisture or Humidity Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21737961 ↗
  • MalfunctionMar 21, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21662155 ↗
  • MalfunctionFeb 21, 2025Gas Output Problem; Gas Output Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21433248 ↗
  • MalfunctionFeb 21, 2025Gas Output Problem; Gas Output Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21432592 ↗
  • MalfunctionFeb 18, 2025Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS FOUND DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21395975 ↗
  • MalfunctionJan 7, 2025Gas Output Problem; Gas Output Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S DRAINAGE VALVE WAS DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21091253 ↗
  • MalfunctionDec 27, 2024Gas Output Problem; Moisture or Humidity Problem; Gas Output Problem; Moisture or Humidity Problem

    IT WAS REPORTED THAT THE COMPRESSOR'S AIR OUTLET WAS CONTAMINATED WITH WATER. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21027498 ↗
  • MalfunctionDec 13, 2024Failure to Power Up; Moisture or Humidity Problem; Failure to Power Up; Moisture or Humidity Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20923582 ↗
  • MalfunctionNov 7, 2024Moisture or Humidity Problem; Output Problem; Moisture or Humidity Problem; Output Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20628925 ↗
  • MalfunctionOct 4, 2024Gas Output Problem; Gas Output Problem

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20369389 ↗

FDA clearance and classification

The FDA regulates product code BTI (compressor, air, portable) under 21 CFR 868.6250 as a Class II device. 21 CFR 868.6250 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Compressor Mini 115V 60Hz 6481779

Has the Compressor Mini 115V 60Hz 6481779 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07325710001653), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Compressor Mini 115V 60Hz 6481779?

The FDA's MAUDE database lists 108 adverse event reports for this device, received between Jan 5, 2023 and May 12, 2026. In those reports, the most common specific device problems are “Output Problem” (37), “Failure to Power Up” (24) and “Moisture or Humidity Problem” (19). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Compressor Mini 115V 60Hz 6481779?

None of the reports are classified as deaths, 0 as injuries and 108 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Compressor Mini 115V 60Hz 6481779 FDA approved?

Devices of this type (product code BTI) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Maquet Critical Care AB.

Who makes the Compressor Mini 115V 60Hz 6481779, and is it still sold?

The FDA's device database lists Maquet Critical Care AB as the company responsible for it (brand Compressor Mini 115V 60Hz, model 6481779). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Compressor, Air, Portable devices

All Compressor, Air, Portable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page