Servo-air Lite 6893200Recalls & reported problems

Maquet Critical Care AB · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Maquet Servo-air Lite ventilator is a floor-standing machine that provides breathing support in hospitals. The 39 reports mostly describe the unit failing its own pre-use self-checks and generating error codes or alarms, along with flow and turbine problems, with no patient harm mentioned. There are 39 malfunctions, no deaths and no injuries. No recalls name this device, and reports rose from 14 to 24 over the past 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 07325710015117), received between Sep 27, 2024 and Sep 29, 2026 — classified as 37 malfunctions.

The most frequently reported specific device problem is “Failure of Device to Self-Test” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 4 reportsMay 2026: 5 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure of Device to Self-Test24
  • Output Problem7
  • Infusion or Flow Problem6
  • Improper Flow or Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions37

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Failure of Device to Self-Test

    It was reported that the ventilator failed the Pressure transducer test during the Pre-use Check. There was no patient involvement. Manufacturer's Ref. #: (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932413 ↗
  • MalfunctionJun 24, 2026Output Problem

    IT WAS REPORTED THAT THE VENTILATOR GENERATED A TECHNICAL ERROR CODE INDICATING TURBINE MODULE COMMUNICATION ERROR. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25606173 ↗
  • MalfunctionJun 17, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REFERENCE #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25530379 ↗
  • MalfunctionJun 2, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25359404 ↗
  • MalfunctionMay 24, 2026Output Problem

    IT WAS REPORTED THAT THE VENTILATOR HAD A TURBINE FAILURE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25266734 ↗
  • MalfunctionMay 20, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25229095 ↗
  • MalfunctionMay 19, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25213995 ↗
  • MalfunctionMay 19, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25213994 ↗
  • MalfunctionMay 14, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT VENTILATOR GENERATED AN ALARM OF LOW O2 CONCENTRATION. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25174931 ↗
  • MalfunctionApr 29, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT THE VENTILATOR HAD ISSUES WITH DELIVERY OF AIR FLOW. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25024466 ↗
  • MalfunctionApr 17, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24908745 ↗
  • MalfunctionApr 14, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24870776 ↗
  • MalfunctionApr 9, 2026Improper Flow or Infusion

    IT WAS REPORTED THAT THE VENTILATOR GENERATED ALARMS INDICATING A O2 CONCENTRATION LOW, CHECK TUBING AND LEAKAGE TOO HIGH. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24827692 ↗
  • MalfunctionMar 23, 2026Failure of Device to Self-Test

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663228 ↗
  • MalfunctionFeb 27, 2026Failure of Device to Self-Test

    MANUFACTURER'S REF #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24469702 ↗
  • MalfunctionFeb 27, 2026Failure of Device to Self-Test

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24468672 ↗
  • MalfunctionFeb 27, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24468601 ↗
  • MalfunctionFeb 16, 2026Failure of Device to Self-Test

    MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24344172 ↗
  • MalfunctionJan 27, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24187806 ↗
  • MalfunctionJan 23, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24159304 ↗
  • MalfunctionJan 21, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24127589 ↗
  • MalfunctionJan 20, 2026Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24113089 ↗
  • MalfunctionOct 31, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23434559 ↗
  • MalfunctionOct 31, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT THE VENTILATOR ALARMED FOR TURBINE FAILURE. THERE WAS NO PATIENT HARM REPORTED. MANUFACTURER'S REF #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23434490 ↗
  • MalfunctionOct 29, 2025Failure of Device to Self-Test; Failure of Device to Self-Test

    IT WAS REPORTED THAT THE VENTILATOR FAILED FLOW TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23411562 ↗

Other Ventilator, Continuous, Minimal Ventilatory Support, Facility Use devices