AUTO EFFLUENT SET 115370Recalls & reported problems

Gambro Uf Solutions, Inc · The Auto Effluent Accessory is indicated for use only with the PrisMax control unit in providing automatic effluent draining during Continuous Renal Replacement Therapy (CRRT). · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07332414118914), received between Mar 8, 2023 and May 24, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Protective Measures Problem1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT WHILE PERFORMING CONTINUOUS RENAL REPLACEMENT THERAPY, A PRISMAX AUTO EFFLUENT BAG WAS LEAKING AT THE JUNCTION OF THE BAG AND THE TUBING. TO RESOLVE THIS ISSUE, THE EFFLUENT BAG WAS REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19397871 ↗
  • MalfunctionMar 8, 2023Fluid/Blood Leak

    THE PATIENT WAS RECEIVING CONTINUOUS RENAL REPLACEMENT THERAPY WITH BAXTER HEALTHCARE PRISMAX MACHINE. THE RN NOTICED EFFLUENT LEAKING FROM THE SMALL 1-LITER BAG OF THE AUTO EFFLUENT ACCESSORY KIT. IMMEDIATELY, THE RN STOPPED THERAPY AND REPLACED THE CRRT TUBING (PRISMAFLEX EXTRA CORPOREAL CIRCUIT WITH FILTER, THERMAX BLOOD WARMER AND AUTO EFFLUENT ACCESSORY KIT. THE REPLACEMENT PROCESS DELAYS THERAPY FOR AT LEAST 20 MINUTES. THE FLUID WAS LEAKING FROM THE JUNCTION OF TUBING AND BAG (PICTURES AVAILABLE). THE BAXTER REP WILL BE REMOVING ALL AUTO EFFLUENT ACCESSORY PRODUCTS WITH LOT # 22H2901 AND REPLACING WITH NEW PRODUCTS.

    Patient problems: Insufficient Information

    MAUDE report 16501519 ↗
  • MalfunctionMar 8, 2023Protective Measures Problem

    THE PATIENT WAS RECEIVING CRRT, AND THE RN NOTICED EFFLUENT DRIPPING FROM THE 1-LITER AND 5-LITER BAGS. THE AUTO-EFFLUENT DRAIN ACCESSORY CONTINUOUSLY PUMPS EFFLUENT COLLECTED DURING THERAPY INTO A DRAIN, WHILE PROVIDING THE ACCURACY OF A SCALE-BASED MEASUREMENT SYSTEM. ELIMINATING EFFLUENT BAG CHANGES MINIMIZES USER EXPOSURE TO EFFLUENT. THE LOSS OF EFFLUENT CAN AFFECT THE PATIENT'S THERAPY AND INCREASE EXPOSURE OF HEALTHCARE WORKER TO EFFLUENT. THE RN HAD TO COMPLETELY REMOVE THE CURRENT CRRT SET PRISMAFLEX EXTRA CORPOREAL CIRCUIT WITH FILTER, THERMAX BLOOD WARMER, AND AUTO EFFLUENT DRAIN ACCESSORY. THE MACHINE WOULD NOT ALLOW THE RN TO ONLY REPLACE THE AUTO EFFLUENT DRAIN ACCESSORY. CHANGING OF THE TUBING CAUSES A DELAY IN THERAPY FOR AT LEAST 20 MINUTES. ALSO, THE COST INCREASES BECAUSE THE CRITICAL CARE UNIT HAS TO UTILIZE ANOTHER SET EARLY. THE BAXTER REP TOLD THE LEADERSHIP TO REMOVE ALL SETS WITH THE LOT # 22H2901. THE REP WILL TAKE THE TUBING WITH LOT # 22H2901 AND REPLACE WITH A NEW LOT #.

    Patient problems: Insufficient Information

    MAUDE report 16501478 ↗

Other SYSTEM, HEMODIALYSIS, REMOTE ACCESSORIES devices

All SYSTEM, HEMODIALYSIS, REMOTE ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page