REVACLEAR 300 (REF 114745L)Recalls & reported problems

Gambro Renal Products, Inc. · Revaclear 300 and 400 dialyzers are indicated for the treatment of chronic and acute renal failure by hemodialysis · Class 2

All models of REVACLEAR (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Revaclear 300 dialyzer is a filter used in hemodialysis for kidney failure. Most of the 125 reports describe fluid or blood leaks from the dialyzer, and some describe contamination with particles. The reports include no deaths and 13 injuries. There are 3 recalls, 1 open, including one over material shifting inside the dialyzer that can look like a blood leak.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
125
Deaths reported
0
Injuries reported
13
Recalls naming it
3

Summary

The FDA's MAUDE database contains 125 adverse event reports that list this exact device (DI 07332414123055), received between Jan 19, 2018 and Jan 6, 2025 — classified as 13 injuries, 112 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (71 reports).

3 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 23, 2022Class IIOpen, Classified

    Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak

    Z-0762-2023 · Root cause: Process control · Recall details · FDA record ↗

  2. Nov 15, 2019Class IITerminated

    There is the potential presence of particular matter in the header caps of ten lots the Revaclear 300 Dialyzer.

    Z-1409-2020 · Root cause: Process control · Recall details · FDA record ↗

  3. May 31, 2019Class IITerminated

    There is a potential presence of ruptured dialyzer fibers which may lead to a blood leak during treatment.

    Z-1908-2019 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak71
  • Leak/Splash25
  • Device Contamination with Chemical or Other Material19
  • Adverse Event Without Identified Device or Use Problem9
  • Break1
  • Crack1
  • Material Twisted/Bent1

Patient problems

  • No Consequences Or Impact To Patient71
  • No Clinical Signs, Symptoms or Conditions25
  • No Patient Involvement13
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Dyspnea4
  • Hemolysis4
  • Low Blood Pressure/ Hypotension3
  • Malaise2

Recent adverse event reports

The 25 most recent of 125 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM AN UNSPECIFIED LOCATION OF A REVACLEAR 300 DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21084613 ↗
  • MalfunctionJun 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE END CAP OF A REVALCEAR 300 DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19601527 ↗
  • InjuryJun 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THREE MINUTES AFTER STARTING TREATMENT WITH A REVACLEAR 300 DIALYZER, THE PATIENT EXPERIENCED BACK PAIN, ITCHING, NAUSEA AND VOMITING, COLD/CLAMMY SKIN AND PALLOR, LIGHTHEADED, SHORTNESS OF BREATH AND DECREASED SYSTOLIC BLOOD PRESSURE TO "90S" . THE TREATMENT WAS DISCONTINUED,AND THE DIALYZER WAS REPLACED WITH A NON-BAXTER DIALYZER. THE PATIENT WAS GIVEN BENADRYL AND NORMAL SALINE AS TREATMENT. THE PATIENT¿S SYMPTOMS HAD "MARGINAL IMPROVEMENT". DIALYSIS TREATMENT WAS TERMINATED 2 HOURS AND 45 MINUTES EARLY AND THE PATIENT WAS SENT TO THE EMERGENCY ROOM VIA AMBULANCE. THE PATIENT WAS NOT ADMITTED TO THE HOSPITAL. THE PATIENT RETURNED TO THE DIALYSIS CLINIC THE FOLLOWING DAY TO "MAKE UP THE TREATMENT". NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Dizziness; Dyspnea; Hypersensitivity/Allergic reaction; Itching Sensation; Low Blood Pressure/ Hypotension; Nausea; Pain; Pallor; Vomiting

    MAUDE report 19538843 ↗
  • MalfunctionSep 19, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED UNSPECIFIED PARTICULATE WAS OBSERVED IN SIXTEEN (16) STERILIZED REVACLEAR 300 SETS. THE PARTICULATE WAS NOTED DURING AN UNKNOWN PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17773219 ↗
  • MalfunctionSep 19, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED UNSPECIFIED PARTICULATE WAS OBSERVED IN SIXTEEN (16) STERILIZED REVACLEAR 300 SETS. THE PARTICULATE WAS NOTED DURING AN UNKNOWN PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17773180 ↗
  • InjurySep 7, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT APPROXIMATELY THREE MINUTES AFTER STARTING TREATMENT WITH REVACLEAR 300, A BLOOD LEAK ALARM WAS GENERATED. BLOOD WAS OBSERVED IN THE DIALYSATE PATH. TREATMENT WAS STOPPED. IT WAS FURTHER REPORTED THE BLOOD WAS NOT RETURNED TO THE PATIENT AND A BLOOD TRANSFUSION WAS ADMINISTERED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17697758 ↗
  • MalfunctionJun 26, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AN EXTERNAL BLOOD LEAK WAS OBSERVED ORIGINATING FROM THE ¿TOP OF THE DIALYZER ON THE OUTSIDE¿ OF AN UNSPECIFIED QUANTITY OF REVACLEAR 300 SETS DURING HEMODIALYSIS THERAPY. THE VOLUME OF BLOOD LOST WAS NOT REPORTED FOR ANY OF THE EVENTS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17199404 ↗
  • MalfunctionJun 26, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AN EXTERNAL BLOOD LEAK WAS OBSERVED ORIGINATING FROM THE ¿TOP OF THE DIALYZER ON THE OUTSIDE¿ OF A REVACLEAR 300 SET DURING HEMODIALYSIS THERAPY. THE VOLUME OF BLOOD LOST WAS NOT REPORTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17199402 ↗
  • MalfunctionJun 26, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AN EXTERNAL BLOOD LEAK WAS OBSERVED ORIGINATING FROM THE ¿TOP OF THE DIALYZER ON THE OUTSIDE¿ OF A REVACLEAR 300 SET DURING HEMODIALYSIS THERAPY. THE VOLUME OF BLOOD LOST WAS NOT REPORTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17199401 ↗
  • MalfunctionMay 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AN EXTERNAL BLOOD LEAK WAS OBSERVED ORIGINATING FROM AT THE FILTER'S VENOUS CONNECTOR OF A REVACLEAR 300 APPROXIMATELY ¿FIVE MINUTES¿ INTO HEMODIALYSIS THERAPY. THE BLOOD LEAK ALARM WAS TRIGGERED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16994937 ↗
  • MalfunctionMay 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AN EXTERNAL BLOOD LEAK WAS OBSERVED ORIGINATING FROM AN UNSPECIFIED LOCATION OF A REVACLEAR 300 DURING HEMODIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16993780 ↗
  • MalfunctionMay 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED TWO (2) EXTERNAL BLOOD LEAKS WERE OBSERVED ORIGINATING FROM AN UNSPECIFIED LOCATION OF A REVACLEAR 300 DURING HEMODIALYSIS THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16993441 ↗
  • InjuryMay 22, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING TREATMENT WITH A REVACLEAR 300, THE MACHINE TRIGGERED AN INTERNAL BLOOD LEAK ALARM, TREATMENT WAS DISCONTINUED AND THE EXTRACORPOREAL CIRCUIT WAS NOT RETURNED TO THE PATIENT. THE PATIENT PRESENTED TO THE HOSPITAL DUE TO AN ESTIMATED BLOOD LOSS OF 900ML. THE PATIENT EXPERIENCED SHORTNESS OF BREATH AND AN OVERALL FEELING OF MALAISE. OXYGEN WAS ADMINISTERED TO THE PATIENT. THE PATIENT RECEIVED DIALYSIS AT THE HOSPITAL IN AN ACUTE SETTING WITH NO ISSUES. THE PATIENT WAS DISCHARGED THE SAME DAY. THE PATIENT OUTCOME WAS REPORTED AS "GOOD". THE PATIENT HAS SINCE DIALYZED PER THEIR NORMAL SCHEDULE IN THE CLINIC WITH NO ISSUES NOTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Dyspnea; Malaise

    MAUDE report 16981263 ↗
  • InjuryMay 22, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AFTER STARTING TREATMENT WITH A REVACLEAR 300, THE DIALYZER TURNED VERY DARK, "ALMOST BLACK" AFTER THE BLOOD HIT THE DIALYZER. THE BLOOD RETURN WAS STARTED; HOWEVER ; THEN STOPPED DUE TO THE PATIENT COMPLAINING OF SHORTNESS OF BREATH AND RESTLESSNESS. THE EXTRACORPOREAL CIRCUIT WAS NOT RETURNED TO THE PATIENT. OXYGEN WAS ADMINISTERED TO THE PATIENT, AND THE PATIENT REPORTED THEY FELT "MUCH BETTER". TREATMENT WAS THEN RESTARTED WITH A DIFFERENT REVACLEAR 300 DIALYZER FROM SAME LOT, HOWEVER, THE MACHINE TRIGGERED AN INTERNAL BLOOD LEAK ALARM, AND THE EXTRACORPOREAL CIRCUIT WAS NOT RETURNED TO THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Anxiety; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Dyspnea

    MAUDE report 16975610 ↗
  • InjuryApr 17, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED TWO (2) INTERNAL BLOOD LEAKS WERE OBSERVED IN A REVACLEAR 300. FORTY MINUTES INTO HEMODIALYSIS THERAPY, THE UNSPECIFIED HEMODIALYSIS MACHINE ALARMED INTERNAL BLOOD LEAK. THE PATIENT LOST 250ML OF BLOOD. THE PATIENT WAS RESTARTED ON A NEW MACHINE AND NEW REVACLEAR (FROM THE SAME LOT). AGAIN 40 MINUTES INTO THERAPY, THE MACHINE ALARMED INTERNAL BLOOD LEAK. THE PATIENT LOST ANOTHER 250ML OF BLOOD. THE PATIENT WAS SET UP WITH A NEW MACHINE AND A REVACLEAR FROM A DIFFERENT LOT NUMBER. THERAPY WAS COMPLETED WITH NO ISSUES. THE FOLLOWING DAY, THE PATIENT UNDERWENT A BLOOD TRANSFUSION. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 16755799 ↗
  • MalfunctionMar 28, 2023Material Twisted/Bent

    NEW DIALYZER OPENED TO PLACE ON THE MACHINE. UPON INSPECTION, THE FIBERS WERE BENT AND LOOK AS IF THEY ARE NOT CONNECTED AS THEY SHOULD BE. DIALYZER PULLED FROM USE.

    Patient problems: Insufficient Information

    MAUDE report 16627223 ↗
  • MalfunctionJan 6, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING USE WITH A REVACLEAR, AN EXTERNAL BLOOD LEAK WAS OBSERVED. THE BLOOD WAS NOT RETURNED TO THE PATIENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16111800 ↗
  • MalfunctionJan 5, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING THE PRIMING OF A REVACLEAR 300, A SALINE SOLUTION LEAKAGE THROUGH THE FIBERS WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16102944 ↗
  • MalfunctionDec 28, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING THE PRIMING OF A REVACLEAR 300 , A SALINE SOLUTION LEAKAGE WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16059598 ↗
  • MalfunctionDec 20, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING PRIMING WITH A REVACLEAR 300, A SALINE SOLUTION LEAK WAS OBSERVED THROUGH ONE OF THE ENDS OF THE FILTER. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16014971 ↗
  • MalfunctionNov 22, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING TREATMENT WITH ONE UNIT OF REVACLEAR 300 , AN EXTERNAL BLOOD LEAK WAS OBSERVED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15841335 ↗
  • MalfunctionSep 28, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AFTER FIVE MINUTES OF TREATMENT WITH ONE UNIT OF REVACLEAR 300, AN EXTERNAL FLUID LEAKAGE WAS NOTED FROM THE TOP OF THE DIALYZER PORT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15496615 ↗
  • MalfunctionAug 2, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING PRIMING WITH A REVACLEAR 300, AN EXTERNAL FLUID LEAK "FROM ONE OF THE ENDS" WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15147720 ↗
  • MalfunctionApr 11, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT HAIR WAS OBSERVED UPON OPENING ONE UNIT OF REVACLEAR 300. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14075124 ↗
  • MalfunctionFeb 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING PRIMING WITH A REVACLEAR 300, AN EXTERNAL DIALYSATE LEAKAGE AT THE LEVEL OF THE VENOUS POLE WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13518077 ↗

FDA clearance and classification

The FDA regulates product code KDI (dialyzer, high permeability with or without sealed dialysate system) under 21 CFR 876.5860 as a Class II device. 21 CFR 876.5860 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Common questions about the REVACLEAR 300

Has the REVACLEAR 300 been recalled?

Yes. 3 FDA recall records name this device's identifier: 1 still open and 2 closed (terminated) by the FDA. The most recent, Z-0762-2023 (Class II), began on Nov 23, 2022: Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the REVACLEAR 300?

The FDA's MAUDE database lists 125 adverse event reports for this device, received between Jan 19, 2018 and Jan 6, 2025. In those reports, the most common specific device problems are “Fluid/Blood Leak” (71), “Leak/Splash” (25) and “Device Contamination with Chemical or Other Material” (19); the most common patient problems are “Dyspnea” (4), “Hemolysis” (4) and “Low Blood Pressure/ Hypotension” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the REVACLEAR 300?

None of the reports are classified as deaths, 13 as injuries and 112 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the REVACLEAR 300, and is it still sold?

The FDA's device database lists Gambro Renal Products, Inc. as the company responsible for it (brand REVACLEAR, model 300). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Dialyzer, high permeability with or without sealed dialysate system devices

All Dialyzer, high permeability with or without sealed dialysate system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page