Skip to content

Brånemark System® 28898Recalls & reported problems

Nobel Biocare AB · Brånemark System Mk III WP 5.0x8.5mm · Class 2 · Implant

All models of Brånemark System (141)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07332747004823), received between Sep 15, 2019 and Jul 24, 2025 — classified as 10 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration8
  • Failure to Osseointegrate2

Patient problems

  • Failure of Implant8
  • Unspecified Infection2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 24, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Infection

    MAUDE report 22595408 ↗
  • InjuryJul 24, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Infection

    MAUDE report 22595403 ↗
  • InjuryDec 20, 2023Loss of Osseointegration; Loss of Osseointegration

    IMPLANT FAILED DUE TO A LOSS OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 18368283 ↗
  • InjuryDec 20, 2023Loss of Osseointegration; Loss of Osseointegration

    IMPLANT FAILED DUE TO A LOSS OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 18368210 ↗
  • InjuryAug 25, 2023Loss of Osseointegration; Loss of Osseointegration

    IMPLANT FAILED DUE TO A LOSS OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 17628467 ↗
  • InjuryAug 25, 2023Loss of Osseointegration; Loss of Osseointegration

    IMPLANT FAILED DUE TO A LOSS OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 17627768 ↗
  • InjuryAug 29, 2022Loss of Osseointegration; Loss of Osseointegration

    IMPLANT FAILED DUE TO A LOSS OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 15315539 ↗
  • InjuryAug 29, 2022Loss of Osseointegration; Loss of Osseointegration

    IMPLANT FAILED DUE TO A LOSS OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 15315484 ↗
  • InjurySep 15, 2019Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILED DUE TO FAILURE TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 9067631 ↗
  • InjurySep 15, 2019Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILED DUE TO FAILURE TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 9063998 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page