Xpert GBS LB GXGBS-LB-10Recalls & reported problems
Cepheid · Xpert GBS LB kit; 10 test kit · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07332940000882), received between Jul 12, 2023 and Aug 30, 2023 — classified as 2 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Septic Shock2
- Convulsion/Seizure1
- Feeding Problem1
- Meningitis1
- Pulmonary Dysfunction1
- Respiratory Distress Syndrome of Newborns1
- Unspecified Respiratory Problem1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
US CUSTOMER CONTACTED CEPHEID TO DISCUSS DISCREPANT RESULTS WITH XPERT GBS LB V1 ASSAY. ON (B)(6) 2023, PATIENT 1 SAMPLE 1 GENITAL SAMPLE WAS COLLECTED FOR ROUTINE GBS TESTING BEFORE DELIVERING HER CHILD. THE SAMPLE WAS RECEIVED BY THE LAB, PLACED INTO LIM BROTH AND INCUBATED OVERNIGHT. ON (B)(6) 2023, A CLEAN SWAB WAS PLACED INTO BROTH AND THEN PUT INTO THE CARTRIDGE AND RAN ON THE XPERT GBS LB ASSAY. THE RESULTS WERE GBS NEGATIVE. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE LAB NO LONGER HAD THIS SPECIMEN, AND THEY DID NOT DO A CULTURE AT THE TIME OF COLLECTION/TESTING. IT IS UNKNOWN IF THE PATIENT WAS ON ANY MEDICATION AS NOTHING WAS CHARTED, OTHER THAN NOTATING SHE WAS ALLERGIC TO RED DYE. AFTER THE BABY WAS BORN, THEY BEGAN SHOWING SYMPTOMS, AND BOTH A BLOOD SAMPLE AND CSF SAMPLE TURNED POSITIVE FOR GBS RESULTS. THE BABY BECAME CRITICALLY ILL. THESE SAMPLES WERE RUN ON BIOFIRE. ON (B)(6) 2023, PATIENT 1 SAMPLE 2S GENITAL SWAB WAS COLLECTED SO THE TEAM COULD RECOLLECT AND RETEST THE MOTHER. A CULTURE WAS CREATED FROM THIS COLLECTION. ON (B)(6) 2023, PATIENT 1 SAMPLE 2 1ST RUN WAS PROCESSED THE SAME WAY AS ABOVE ON THE XPERT GBS LB V1 ASSAY. THE RESULTS WERE GBS NEGATIVE. ON
Patient problems: Convulsion/Seizure; Meningitis; Respiratory Distress Syndrome of Newborns; Septic Shock
MAUDE report 17653313 ↗InjuryJul 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
US CUSTOMER CONTACTED CEPHEID TO DISCUSS DISCREPANT RESULTS WITH XPERT GBS-LB. ON (B)(6) 2023, THE CUSTOMER COLLECTED A VAGINAL/RECTAL SWAB FROM THE PATIENT. THE CUSTOMER MIXED THE ESWAB WITH INVERSION AND THEN PIPETTED 200UL OF THE ESWAB INTO THE LIM BROTH (REMEL TODD HEWITT W/CNA) AND INCUBATED FOR 18 HOURS. ON (B)(6) 2023, THE CUSTOMER THEN PROCESSED THE LIM BROTH PER PI AND RAN IT ON XPERT GBS-LB, WHICH RESULTED IN GBS NEGATIVE. THIS WAS REPORTED TO THE MD. ON (B)(6) 2023, THE CUSTOMER SUBBED THE ORIGINAL ESWAB TO REMEL SHEEP BLOOD AGAR. ON THE SAME DAY, THE CUSTOMER ALSO DECIDED TO USE THE SAME ESWAB TO INOCULATE A BRAND-NEW LIM BROTH VIAL WITH 200UL OF THE ESWAB. CUSTOMER DID NOT INFORM HOW ORIGINAL SWAB WAS STORED IN BETWEEN INOCULATIONS. DUE TO THE NEWBORN BABY OF THE PATIENT BEING SYMPTOMATIC, THE CUSTOMER DID NOT TRUST THIS RESULT AND DECIDED TO RUN THE ORIGINAL ENRICHED LIM BROTH AGAIN. ON (B)(6) 2023, THE CUSTOMER THEN PROCESSED THE LIM BROTH PER PI AND RAN IT ON XPERT GBS-LB, WHICH RESULTED IN GBS NEGATIVE. THIS WAS NOT REPORTED TO THE MD. THE CUSTOMER THEN DECIDED TO TRY RUNNING THIS SAMPLE AGAIN ON THE ORIGINAL ENRICHED LIM BROTH AGAIN. ON (B)(6) 2023, THE CUSTOMER T
Patient problems: Feeding Problem; Pulmonary Dysfunction; Septic Shock; Unspecified Respiratory Problem
MAUDE report 17313221 ↗
Other Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test devices
- Alethia™ Group B Streptococcus 480350Meridian Bioscience, Inc. · Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test1 recall0 reports
- BD MAX™ GBS 441772Becton, Dickinson And Company · Nucleic acid amplification assay system, group b streptococcus, direct specimen test38 reports
- Xpert GBS GXGBS-100N-10Cepheid · Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test1 report
- illumigene Group B Streptococcus 280350Meridian Bioscience, Inc. · Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page