Skip to content

Leksell Stereotactic System 1002248Recalls & reported problems

Elekta Instrument AB · FRAME G WITH STRAIGHT AND CURVED FRONT · Class 2

All models of Leksell Stereotactic System (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07340048306348), received between Jul 25, 2018 and Feb 1, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Defective Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K152558); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Class IIOpen, Classified

    The incorrect IFU was distributed with the devices.

    Z-1309-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1
  • Insufficient Device Problem Information1
  • Use of Device Problem1

Patient problems

  • Burn(s)1
  • Burn, Thermal1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 1, 2024

    PATIENT UNDERGOING BILATERAL DBS ELECTRODE IMPLANT USING LEKSELL FRAME. ONE ELECTRODE WAS IMPLANTED, AND CT USED TO CHECK PLACEMENT. ERROR WAS GREATER THAN ACCEPTABLE OR EXPECTED. THE SPARE ARC/RING WAS STERILIZED, AND THE ELECTRODE WAS REIMPLANTED. ERROR WAS IMPROVED, BUT WAS NOT ACCEPTABLE AND NOT EXPLAINABLE BY HUMAN ERROR OR NATURAL DEVIATIONS. THE INCISION WAS CLOSED. INTENSIVE TROUBLESHOOTING WAS PERFORMED. MY PARTNER WAS CONSULTED AS WELL FOR ADDITIONAL PERSPECTIVES ON CAUSES. ADDITIONAL SCANNING, AND "PHANTOM" TARGETS THAT WERE NON-INVASIVE WERE TARGETED WITH THE ARC/RING. THE LEKSELL FIDUCIAL BOX WAS TESTED, WITH HIGH BUT NOT UNACCEPTABLE ERROR PER NAVIGATION SYSTEM. OPERATION WAS ABORTED DUE TO LACK OF FUNCTIONING EQUIPMENT; MORE THAN ONE POSSIBLE MALFUNCTIONING PIECE OF EQUIPMENT, WITH A NON-LINEAR ERROR THAT COULD NOT BE RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18630647 ↗
  • MalfunctionNov 15, 2019Use of Device Problem; Use of Device Problem; Defective Device

    THE CUSTOMER REPORTED A LEKSELL SYSTEM INACCURACY.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9326679 ↗
  • MalfunctionJul 25, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT DURING THE MR SCAN WHILST USING THE FRAME PATIENTS HAVE EXPERIENCED BURN MARKS. BASED ON THE AVAILABLE INFORMATION ACTUAL INJURY HAS OCCURRED.

    Patient problems: Burn(s); Burn, Thermal

    MAUDE report 7718229 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K152558. One submission often covers many models, so check the product and lot details in each record.

  • Z-0227-2025 · Sep 25, 2024 · Open, Classified
    Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for
  • Z-1397-2024 · Mar 15, 2024 · Open, Classified
    Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Ki