Leksell Stereotactic System 1002248Recalls & reported problems
Elekta Instrument AB · FRAME G WITH STRAIGHT AND CURVED FRONT · Class 2
All models of Leksell Stereotactic System (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07340048306348), received between Jul 25, 2018 and Feb 1, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K152558); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 1, 2023Class IIOpen, Classified
The incorrect IFU was distributed with the devices.
Z-1309-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Defective Device1
- Insufficient Device Problem Information1
- Use of Device Problem1
Patient problems
- Burn(s)1
- Burn, Thermal1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 1, 2024
PATIENT UNDERGOING BILATERAL DBS ELECTRODE IMPLANT USING LEKSELL FRAME. ONE ELECTRODE WAS IMPLANTED, AND CT USED TO CHECK PLACEMENT. ERROR WAS GREATER THAN ACCEPTABLE OR EXPECTED. THE SPARE ARC/RING WAS STERILIZED, AND THE ELECTRODE WAS REIMPLANTED. ERROR WAS IMPROVED, BUT WAS NOT ACCEPTABLE AND NOT EXPLAINABLE BY HUMAN ERROR OR NATURAL DEVIATIONS. THE INCISION WAS CLOSED. INTENSIVE TROUBLESHOOTING WAS PERFORMED. MY PARTNER WAS CONSULTED AS WELL FOR ADDITIONAL PERSPECTIVES ON CAUSES. ADDITIONAL SCANNING, AND "PHANTOM" TARGETS THAT WERE NON-INVASIVE WERE TARGETED WITH THE ARC/RING. THE LEKSELL FIDUCIAL BOX WAS TESTED, WITH HIGH BUT NOT UNACCEPTABLE ERROR PER NAVIGATION SYSTEM. OPERATION WAS ABORTED DUE TO LACK OF FUNCTIONING EQUIPMENT; MORE THAN ONE POSSIBLE MALFUNCTIONING PIECE OF EQUIPMENT, WITH A NON-LINEAR ERROR THAT COULD NOT BE RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18630647 ↗MalfunctionNov 15, 2019Use of Device Problem; Use of Device Problem; Defective Device
THE CUSTOMER REPORTED A LEKSELL SYSTEM INACCURACY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9326679 ↗MalfunctionJul 25, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT DURING THE MR SCAN WHILST USING THE FRAME PATIENTS HAVE EXPERIENCED BURN MARKS. BASED ON THE AVAILABLE INFORMATION ACTUAL INJURY HAS OCCURRED.
Patient problems: Burn(s); Burn, Thermal
MAUDE report 7718229 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K152558. One submission often covers many models, so check the product and lot details in each record.
- Z-0227-2025 · Sep 25, 2024 · Open, ClassifiedElekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for
- Z-1397-2024 · Mar 15, 2024 · Open, ClassifiedElekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Ki
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page