Leksell Vantage Stereotactic System 1072447Recalls & reported problems

Elekta Instrument AB · LEKSELL VANTAGE FRAME HOLDER · Class 2

All models of Leksell Vantage Stereotactic System (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07340048311410), received between Dec 16, 2019 and Jun 18, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K190887); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Handling Problem1
  • Material Too Soft/Flexible1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2025Device Handling Problem; Device Handling Problem

    THE CUSTOMER REPORTED THAT A NURSE MANAGED TO ATTACHED THE FRAME HOLDER INCORRECTLY RESULTING IN DBS ELECTRODE BEING MISPLACED BY ~1CM. THE POSITION WAS CORRECTED USING A DIFFERENT FRAME HOLDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22282531 ↗
  • MalfunctionDec 16, 2019Material Too Soft/Flexible; Unintended Movement; Material Too Soft/Flexible

    DURING VERIFICATION OF VANTAGE FRAME ADAPTER IT WAS OBSERVED THAT THERE IS A RISK OF SHIFTING THE FRAME VERSUS THE ADAPTER IN X-DIRECTION WHEN HOLDING THE FRAME AND ADAPTER ON THE SIDE. THE INTERFACE IS SIMILAR IN THE VANTAGE FRAME HOLDER AND THERE IS A RISK THAT THE SAME OR SIMILAR SHIFT CAN OCCUR DURING HANDLING.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 9473261 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190887. One submission often covers many models, so check the product and lot details in each record.

  • Z-2006-2026 · Mar 19, 2026 · Open, Classified
    Brand Name: Leksell Vantage Arc System REF: 1053958
  • Z-0433-2026 · Sep 30, 2025 · Open, Classified
    Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The Intended Purpose of Leksell Vantage" Ste
  • Z-1826-2020 · Mar 30, 2020 · Terminated
    Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic Instrument