Skip to content

Leksell GammaPlan 1551287Recalls & reported problems

Elekta Instrument AB · Leksell GammaPlan · Class 2

All models of Leksell GammaPlan (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07340048311878), received between Feb 5, 2025 and Feb 5, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2025Open, Classified

    Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

    Z-1340-2026 · Root cause: Use error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 5, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    THE CUSTOMER CONTACTED ELEKTA TO ASK WHETHER THE "SHOT COORDINATES" IN THE GK TREATMENT PLAN REPORT REPRESENT ABSOLUTE COUCH COORDINATES (IN THE MACHINE'S FRAME OF REFERENCE) OR IF THEY VARY DEPENDING ON WHICH SCAN IS USED AS THE STEREOTACTIC REFERENCE.

    Patient problems: Insufficient Information

    MAUDE report 21311042 ↗

Other System, planning, radiation therapy treatment devices

All System, planning, radiation therapy treatment devices →