Sonesta 6210Recalls & reported problems
Sonesta Medical AB · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07350006950018), received between Jan 9, 2026 and Jan 9, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Missing Information” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 22, 2025Class IIOpen, Classified
Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.
Z-0460-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Missing Information1
Patient problems
- Back pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 9, 2026Missing Information
DURING POST-PROCEDURE PATIENT CARE, A PARAPLEGIC PATIENT WAS REPOSITIONED WHILE SEATED ON A MODEL 6210 PROCEDURE CHAIR. DURING DRESSING, THE PATIENT WAS LEANED AGAINST A DETACHABLE ARM REST. THE ARM REST IS INTENDED FOR PATIENT COMFORT AND ARM POSITIONING AND IS NOT DESIGNED TO SUPPORT PATIENT BODY WEIGHT. DURING THIS MANEUVER, THE ARM REST ATTACHMENT CAME LOOSE, AND THE PATIENT FELL FROM THE CHAIR TO THE FLOOR. TWO HEALTHCARE STAFF MEMBERS SUSTAINED BACK INJURIES WHILE ATTEMPTING TO PREVENT THE FALL. ONE STAFF INJURY REQUIRED MEDICAL CARE AND RESULTED IN WORK RESTRICTIONS. THE PATIENT WAS EVALUATED FOLLOWING THE FALL AND NO ACUTE INJURY WAS REPORTED.
Patient problems: Back pain
MAUDE report 24018512 ↗
FDA clearance and classification
The FDA regulates product code IXR (table, radiographic, tilting) under 21 CFR 892.1980 as a Class II device. 21 CFR 892.1980 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Sonesta 6210
Has the Sonesta 6210 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0460-2026 (Class II), began on Aug 22, 2025: Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Sonesta 6210?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 9, 2026. That report lists device problem “Missing Information” (1); patient problem “Back pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Sonesta 6210?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Sonesta 6210 FDA approved?
Devices of this type (product code IXR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Sonesta Medical AB.
Is the Sonesta 6210 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the Sonesta 6210, and is it still sold?
The FDA's device database lists Sonesta Medical AB as the company responsible for it (brand Sonesta, model 6210). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Table, Radiographic, Tilting devices
- EasyDiagnost Eleva 9890-010-03632Philips Medical Systems DMC GmbH · Table, radiographic, tilting1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page