CERAMENT™BONE VOID FILLER A0210-11Recalls & reported problems
Bone Support AB · CERAMENT BONE VOID FILLER is a fast-setting, injectable and moldable ceramic bone graft substitute intended for filling bone voids/gaps. The material consists of a powder and a liquid component. The major constituents of the powder are hydroxyapatite and calcium sulfate hemihydrate. The liquid component (C-TRU) contains Iohexol as a radio-contrast enhancer. Mixing the components with the combined mixing injection (CMI) device, results in a viscous material intended to set ex vivo or in vivo. By combining hydroxyapatite and calcium sulfate a balance is achieved between implant resorption rate and bone in-growth rate. Calcium sulfate acts as a resorbable carrier for hydroxyapatite. Hydroxyapatite has a slow resorption rate, high osteoconductivity, promoting bone in-growth and gives long term structural support to newly formed bone. The ceramic bone graft substitute is placed into the bone defect under visual inspection or under radiographic monitoring during open or percutaneous surgery. The paste may be injected into the defect, molded by hand and digitally placed into the defect, or used to prepare beads that are placed into the defect. The accompanying injection device (ID) and Tip Extenders may be used to facilitate the filling of the bone defect. When fully set in vivo, CERAMENT BONE VOID FILLER is drillable and can be used to augment hardware during the surgical procedure. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07350055430066), received between Jan 4, 2022 and Apr 26, 2024 — classified as 1 death, 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device Dislodged or Dislocated2
- Migration2
- Insufficient Device Problem Information1
Patient problems
- Infiltration into Tissue2
- Inflammation2
- Wound Dehiscence2
- Asystole1
- Bradycardia1
- Thrombosis/Thrombus1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 26, 2024Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
POST-OPERATIVELY THE INCISION HEALED NICELY THEN WEEKS LATER THE INCISION OPENED AND STARTED DRAINING LIQUID CERAMENT. ON SECOND OPERATION, ALL OF THE CERAMENT WITHIN THE CAVITY WAS LIQUID - NOT SOLID. RETURN TO OPERATING ROOM TO REMOVE CERAMENT AND RE-CLOSE THE INCISION. UNNECESSARY ANESTHESIA, COST OF RE-OPERATION. SIGNIFICANT DELAY IN PATIENT'S ABILITY TO RETURN TO ACTIVITY.
Patient problems: Infiltration into Tissue; Inflammation; Wound Dehiscence
MAUDE report 19189466 ↗InjuryApr 26, 2024Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
POST-OPERATIVELY THE INCISION HEALED NICELY THEN WEEKS LATER THE INCISION OPENED AND STARTED DRAINING LIQUID CERAMENT. ON SECOND OPERATION, ALL OF THE CERAMENT WITHIN THE CAVITY WAS LIQUID - NOT SOLID. RETURN TO OPERATING ROOM TO REMOVE CERAMENT AND RE-CLOSE THE INCISION. UNNECESSARY ANESTHESIA, COST OF RE-OPERATION. SIGNIFICANT DELAY IN PATIENT'S ABILITY TO RETURN TO ACTIVITY.
Patient problems: Infiltration into Tissue; Inflammation; Wound Dehiscence
MAUDE report 19189341 ↗InjuryMay 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION GIVEN BY REPORTING SURGEON IN INTERVIEW AND FOLLOW-UP OF INTERVIEW 2023-04-18: ON APRIL 10TH 2023 A MINIMALLY INVASIVE/ PERCUTANEOUS PROCEDURE WAS PERFORMED ON A 67 YEAR OLD MALE PATIENT WITH A DIAGNOSIS OF PLANTAR FASCIITIS AND CONCOMITANT BONE MARROW LESIONS VISIBLE ON PRE-OPERATIVE DIAGNOSTIC MRI SCAN WITHIN THE CALCANEUM. THE PROCEDURE INVOLVED ADDRESSING THE PLANTAR FASCIA SOFT TISSUE COMPONENT USING TENEX. THE BONY COMPONENT OF THE PROCEDURE WAS CARRIED OUT UNDER FLUOROSCOPIC GUIDANCE WITH A MINI-C-ARM AND BOTH AP AND LATERAL IMAGES WAS OBTAINED TO CONCLUDE THAT THE BONESUPPORT DELIVERY CANNULA USED FOR INJECTING THE CERAMENT BVF WAS INDEED IN THE CANCELLOUS BONE OF THE CALCANEUM. THE KIT USED WAS AN 18 CC (ML) VOLUME OF CERAMENT BVF OF WHICH 1.5-2 ML WAS FINALLY USED. IT WAS PROCEEDED TO FILL THE BONE MARROW LESIONS WITHOUT EXERTING TOO MUCH PRESSURE WHILST DOING SO. THERE WAS AN INITIAL RADIO-OPACIFICATION WITHIN THE CALCANEUM RELATED TO THE IOHEXOL COMPONENT OF THE PRODUCT BUT THEN IT BECAME EVIDENT ON FLUOROSCOPY THAT THIS SAME RADIO-OPAQUE MATERIAL WAS MIGRATING AND APPEARED TO HAVE TRACKED INTO A VEIN ON THE LATERAL SIDE. WHEN THIS PHENOMENON WAS RECOGNIZED
Patient problems: Thrombosis/Thrombus
MAUDE report 16872521 ↗DeathJan 4, 2022Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ACCORDING TO THE REPORTER THE FOLLOWING OCCURRED: PATIENT UNDERGOING HIP REVISION EXPERIENCED CARDIAC COLLAPSE WITH DEATH FOLLOWING INJECTION OF CERAMENT BONE VOID FILLER. MICROSCOPIC AUTOPSY EXAMINATION CONCLUDE CAUSE OF DEATH AS EXTENSIVE PULMONARY MICRO-EMBOLI OF FOREIGN MATERIAL CONSISTENT WITH CERAMENT BONE VOID FILLER MATERIAL. NO FAT EMBOLI WERE IDENTIFIED.
Patient problems: Asystole; Bradycardia
MAUDE report 13153147 ↗
Other Filler, Bone Void, Calcium Compound devices
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44150Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND3 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42275Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42280Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42200Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44145Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- MAGNIFUSE® Bone Graft 7509014Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43102Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- HydroSet XT 897005Stryker Corporation · FILLER, BONE VOID, CALCIUM COMPOUND1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page