Skip to content

DS PrimeTaper 68017255Recalls & reported problems

Dentsply Implants Manufacturing GmbH · PrimeTaper Drill Sleeve 3 ND GS · Class 1

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07392532269936), received between Apr 29, 2026 and Sep 29, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam2
  • Detachment of Device or Device Component1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Loose or Intermittent Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938087 ↗
  • MalfunctionJul 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION WHERE THE SLEEVE WAS IMMEDIATELY LOOSE WHEN PLACED ON THE GUIDED DRILL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25676756 ↗
  • MalfunctionApr 29, 2026Mechanical Jam

    IT WAS REPORTED THAT SEVERAL OF THESE SPRINGS HAVE BEEN DISPLACED OR HAVE FRACTURED DURING USE. THIS DOES NOT ALWAYS INVOLVE SIZE 3, BUT IT OCCURS FREQUENTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034223 ↗
  • MalfunctionApr 29, 2026Mechanical Jam

    IT WAS REPORTED THAT SEVERAL OF THESE SPRINGS HAVE BEEN DISPLACED OR HAVE FRACTURED DURING USE. THIS DOES NOT ALWAYS INVOLVE SIZE 3, BUT IT OCCURS FREQUENTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034186 ↗

Other Accessories, implant, dental, endosseous devices

All Accessories, implant, dental, endosseous devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.