Skip to content

DS PrimeTaper 68011106 (UDI 07392532276798)Recalls & reported problems

Dentsply Implants Manufacturing GmbH · PrimeTaper EV Ø4.8 x 11mm OsseoSpeed · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Dentsply PrimeTaper implant has 229 adverse event reports, all of them injuries, and no deaths. The reports mostly describe patients losing the implant, often with infection, bone loss or pain, and some describe no initial stability during surgery. No recalls name this device. Reports rose from 52 in the previous 12 months to 96 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
235
Deaths reported
0
Injuries reported
235
Recalls naming it
0

Summary

The FDA's MAUDE database contains 235 adverse event reports that list this exact device (DI 07392532276798), received between Jun 20, 2023 and Sep 30, 2026 — classified as 235 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (172 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 2 reportsJan 2024: 9 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 5 reportsMay 2024: 7 reportsJun 2024: 6 reportsJul 2024: 5 reportsAug 2024: 6 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 12 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 2 reportsMar 2025: 10 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 3 reportsAug 2025: 4 reportsSep 2025: 6 reportsOct 2025: 6 reportsNov 2025: 6 reportsDec 2025: 8 reportsJan 2026: 11 reportsFeb 2026: 7 reportsMar 2026: 13 reportsApr 2026: 12 reportsMay 2026: 11 reportsJun 2026: 8 reportsJul 2026: 11 reportsAug 2026: 7 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate172
  • Loss of Osseointegration43
  • Fracture9
  • Difficult to Insert5
  • Malposition of Device3
  • Insufficient Device Problem Information2
  • Damaged Thread1
  • Material Deformation1

Patient problems

  • Failure of Implant226
  • Unspecified Infection51
  • Inflammation23
  • Osteolysis20
  • Implant Pain13
  • Pocket Erosion11
  • Inadequate Osseointegration4
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 25 most recent of 235 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Osteolysis

    MAUDE report 50943330 ↗
  • InjurySep 29, 2026Loss of Osseointegration

    It was reported that a patient experienced a dental implant loss.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933480 ↗
  • InjurySep 28, 2026Damaged Thread

    It was reported that a patient experienced a dental implant loss after the insertion tool tip broke off inside the implant.

    Patient problems: Inadequate Osseointegration

    MAUDE report 50917815 ↗
  • InjurySep 17, 2026Fracture

    It was reported that a patient experienced a dental implant breakage.

    Patient problems: Unspecified Infection

    MAUDE report 50805934 ↗
  • InjurySep 15, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 50775069 ↗
  • InjurySep 14, 2026Loss of Osseointegration

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Inadequate Osseointegration

    MAUDE report 50755549 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Inadequate Osseointegration

    MAUDE report 50734370 ↗
  • InjurySep 7, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50651251 ↗
  • InjuryAug 26, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss. Also, it was reported that they had unspecified infection and Osteolysis.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50458276 ↗
  • InjuryAug 25, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss with Periimplantitis.

    Patient problems: Failure of Implant

    MAUDE report 50437782 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Failure of Implant

    MAUDE report 50244328 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Failure of Implant

    MAUDE report 50244293 ↗
  • InjuryAug 12, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Failure of Implant

    MAUDE report 50219882 ↗
  • InjuryAug 9, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss with pain, mobility, and pus.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 50188823 ↗
  • InjuryAug 4, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Failure of Implant

    MAUDE report 50128786 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    It was reported that a patient experienced a dental implant loss.

    Patient problems: Failure of Implant

    MAUDE report 50006296 ↗
  • InjuryJul 14, 2026Failure to Osseointegrate

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS. PATIENT ALSO REPORTED AN INFECTION AND PERIIMPLANTITIS.

    Patient problems: Failure of Implant; Inflammation; Unspecified Infection

    MAUDE report 25816439 ↗
  • InjuryJul 13, 2026Loss of Osseointegration

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.

    Patient problems: Failure of Implant

    MAUDE report 25806517 ↗
  • InjuryJul 13, 2026Failure to Osseointegrate

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS. NOT INTEGRATED

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 25800138 ↗
  • InjuryJul 13, 2026Loss of Osseointegration

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS. ALSO, IT WAS REPORTED THAT THEY HAD UNSPECIFIED INFECTION.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 25794397 ↗
  • InjuryJul 9, 2026Loss of Osseointegration

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.

    Patient problems: Failure of Implant

    MAUDE report 25766127 ↗
  • InjuryJul 9, 2026Loss of Osseointegration

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.

    Patient problems: Failure of Implant

    MAUDE report 25765974 ↗
  • InjuryJul 6, 2026Failure to Osseointegrate

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.

    Patient problems: Failure of Implant

    MAUDE report 25716016 ↗
  • InjuryJul 2, 2026Failure to Osseointegrate

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.

    Patient problems: Failure of Implant

    MAUDE report 25694105 ↗
  • InjuryJul 1, 2026Loss of Osseointegration

    IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT FAILURE.

    Patient problems: Failure of Implant; Osteolysis

    MAUDE report 25685276 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page