TRUWAVE, VAMP PLUS T001691MRecalls & reported problems
Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07460691959477), received between May 7, 2024 and Dec 30, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Backflow” (1 report).
1 other recall cover devices cleared under the same FDA submission (K181684); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Backflow1
- Detachment of Device or Device Component1
- Fluid/Blood Leak1
- Incorrect Measurement1
- Mechanical Jam1
- No Device Output1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2025No Device Output; No Device Output
AS REPORTED, DURING USE IN PATIENT RECEIVING A NITROPRUSSIDE INFUSION, THIS VAMP PRESSURE SET LOST THE BLOOD PRESSURE VALUE ON THE MONITOR, THUS, PUTTING PATIENT AT RISK, AS PER CUSTOMER'S OPINION. REPLACING THE TRANSDUCER SOLVED THE ISSUE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23920565 ↗MalfunctionFeb 6, 2025Incorrect Measurement; Incorrect Measurement
AS REPORTED, DURING USE IN PATIENT, THIS DISPOSABLE PRESSURE TRANSDUCER WITH VAMP PLUS PROVIDED ARTERIAL BLOOD PRESSURE VALUES (200/90) SIGNIFICANTLY DIFFERENT THAN THE ONES PROVIDED BY A NON-INVASIVE DEVICE (160/80). MEASUREMENTS WERE MADE BY BOTH DEVICES IN THE SAME ARM. THE VALUES EXPECTED ACCORDING TO PATIENT STATUS WERE AROUND 140-160/80. NO ERROR MESSAGE NOR ALARM WAS DISPLAYED. THE PATIENT WAS NOT TREATED ACCORDING TO WRONG VALUES PROVIDED. THE DEVICE WAS REPLACED BY A NEW ONE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION SINCE IT WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21322119 ↗MalfunctionJul 2, 2024Mechanical Jam; Mechanical Jam
AS REPORTED, DURING USE IN PATIENT WITH THIS DISPOSABLE PRESSURE TRANSDUCER WITH VAMP PLUS, WHEN THE USER WENT TO TAKE A SAMPLE FOR LABORATORY TESTS, THE DISTAL SAMPLING SITE WAS FOUND COMPLETELY COLLAPSED AND IT WAS NOT POSSIBLE TO BE RETURNED TO ITS INITIAL POSITION. THE BLOOD WAS LEAKING CONTINUOUSLY. THE DEVICE WAS CHANGED FOR A NEW ONE AND NO ADDITIONAL INTERVENTION WAS NEEDED TO SOLVE THE ISSUE. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19655921 ↗MalfunctionMay 17, 2024Backflow; Fluid/Blood Leak; Backflow; Fluid/Blood Leak
AS REPORTED, DURING USE IN PATIENT WITH THIS DISPOSABLE PRESSURE TRANSDUCER WITH VAMP SYSTEMS, THERE WAS A SALINE LEAKAGE AT THE LEVEL OF THE SHUT-OFF VALVE. IN ADDITION, THERE WAS A BLOOD BACKFLOW DUE TO LOSS OF AIRTIGHTNESS OF THE SYSTEM THAT MUST BE CLOSE. THEREFORE THE SYSTEM MUST BE CHANGED BECAUSE IT PRESENTED FILTRATION AND ENTERED AIR INTO THE SYSTEM WITH RISK OF COAGULATION AND DYSFUNCTION. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19339341 ↗MalfunctionMay 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING USE IN PATIENT WITH THIS TRUWAVE VAMP PLUS, THE PRESSURE TUBING DETACHED AND THE SOLUTION LEAKED. THERE WAS NO PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19257676 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K181684. One submission often covers many models, so check the product and lot details in each record.
- Z-0414-2024 · Oct 4, 2023 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)
Other Catheter, Continuous Flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.