Seraclone® Anti-S (MNS3) 808052100Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Blood Grouping Reagent Anti-S (MNS3); Seraclone® Human Monoclonal (MS94); for tube test; 2ml
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07611969951536), received between Apr 26, 2018 and May 27, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- False Negative Result1
- False Positive Result1
Patient problems
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2020False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH ANTI-S WHEN THE PATIENT HAS A POSITIVE DAT. THE CUSTOMER HAS HAD TWO PATIENTS WHERE GENOTYPE TESTING FROM A REFERENCE LAB INDICATED THAT THEY WERE S NEGATIVE AND USING SERACLONE ANTI-S LOT 8913180-01 THE RESULTS WERE POSITIVE 1+. ONE OF THE PATIENT'S CELLS WERE TREATED WITH EGA (CHEMICAL TREATMENT TO DISASSOCIATE IGG FROM THE CELLS) AND THE TESTING WAS REPEATED. THE REACTION WITH SERACLONE ANTI-S WAS STILL POSITIVE, EVEN STRONGER. THE DATE OF EVENT IS NOT KNOWN. THE CUSTOMER SAID TO SEND IN THE PATIENT SAMPLE AND THE SUPPOSEDLY DEFECTIVE PRODUCT SERACLONE ANTI-S FOR INVESTIGATIONAL TESTING. OUR QUALITY CONTROL LABORATORY IS STILL WAITING FOR THE MATERIAL. STILL WE ARE NOT IN THE POSITION TO PROVIDE A CONCLUSIVE STATEMENT. THE ROOT CAUSE ANALYSIS IS STILL ONGOING. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10091543 ↗MalfunctionApr 26, 2018False Negative Result
THE CUSTOMER REPORTED FALSE NEGATIVE REACTIONS OF TWO DONOR SEGMENTS WITH A HISTORY OF ANTI-S WHEN TESTED WITH SERACLONE ANTI-S. THE CUSTOMER DID NOT PROVIDE THE SUPPOSEDLY DEFECTIVE PRODUCT NOR THE DONOR SEGMENTS THAT HAD CAUSED THE FALSE NEGATIVE RESULTS. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THE RETENTION SAMPLE FOR POTENCY AND SPECIFICITY. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE NEGATIVE REACTION. THE MINIMUM TITER WAS EXCEEDED. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THE ALLEGEDLY DEFECTIVE LOT OF SERACLONE ANTI-S FUNCTIONS CORRECTLY. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT.
Patient problems: No Information
MAUDE report 7463642 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page