Seraclone® Anti-s (MNS4) 808068100Recalls & reported problems

Bio-Rad Medical Diagnostics GmbH · Blood Grouping Reagent Anti-s (MNS4); Seraclone® Human Monoclonal (P3YAN3); for tube test; 2ml

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07611969951550), received between Jan 30, 2025 and Jan 30, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2025False Negative Result; False Negative Result

    THE CUSTOMER REPORTED FALSE NEGATIVE RESULTS WITH SERACLONE ANTI-S. THE CUSTOMER PERFORMED A LOT-TO-LOT COMPARISON (PLEASE ALSO SEE REPORT 9610824-2025-00002) BETWEEN TWO LOTS OF SERACLONE ANTI-S ON (B)(6) 2024, USING HETEROZYGOTES PANEL CELLS FOR THEIR POSITIVE QC AND THE REACTION STRENGTH WAS 3+ FOR ALL POSITIVE CELLS TESTED. THESE SAME CELLS WERE TESTED AGAIN ON (B)(6) 2024 AND SHOWED WEAK 1+ REACTIONS. USING A NEW PANEL RECEIVED EARLIER IN THE DAY, THE CELL THAT WAS TESTED WITH BOTH LOTS SHOWED NEGATIVE RESULTS. THE CUSTOMER ANNOUNCED TO SEND IN A PRODUCT SAMPLE OF THE AFFECTED LOTS OF SERACLONE-S. SO FAR, THE SAMPLES HAVE NOT ARRIVED ON OUR PREMISES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21272598 ↗
  • MalfunctionJan 30, 2025False Negative Result; False Negative Result

    THE CUSTOMER REPORTED FALSE NEGATIVE RESULTS WITH SERACLONE ANTI-S. THE CUSTOMER PERFORMED A LOT-TO-LOT COMPARISON (PLASE ALSO SEE REPORT 9610824-2025-00001) BETWEEN TWO LOTS OF SERACLONE ANTI-S ON (B)(6)2024, USING HETEROZYGOTES PANEL CELLS FOR THEIR POSITIVE QC AND THE REACTION STRENGTH WAS 3+ FOR ALL POSITIVE CELLS TESTED. THESE SAME CELLS WERE TESTED AGAIN ON (B)(6)2024 AND SHOWED WEAK 1+ REACTIONS. USING A NEW PANEL RECEIVED EARLIER IN THE DAY, THE CELL THAT WAS TESTED WITH BOTH LOTS SHOWED NEGATIVE RESULTS. THE CUSTOMER ANNOUNCED TO SEND IN A PRODUCT SAMPLE OF THE AFFECTED LOTS OF SERACLONE-S. SO FAR, THE SAMPLES HAVE NOT ARRIVED ON OUR PREMISES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21272506 ↗