Seraclone® Anti-Jkb (JK2) 808184100Recalls & reported problems

Bio-Rad Medical Diagnostics GmbH · Blood Grouping Reagent Anti-Jkb (JK2); Seraclone® Human Monoclonal (MS8); for tube test; 2ml

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07611969951796), received between Dec 23, 2020 and Sep 8, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2025False Negative Result; False Negative Result

    THE CUSTOMER REPORTED A FALSE NEGATIVE REACTION OF SERACLONE ANTI-JKB TO OUR DISTRIBUTOR IN SWITZERLAND. THE CUSTOMER CLAIMS TO HAVE PERFORMED INPUT CONTROL OF THE REAGENT IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE. THE POSITIVE CONTROL, USING TEST ERYTHROCYTES WITH HETEROZYGOUS ANTIGEN EXPRESSION (3% ERYTHROCYTE SUSPENSION BIOTESTCELL P3), SHOWED A NEGATIVE RESULT. THE PROCEDURE WAS REPEATED TWICE, AND BOTH TIMES THE POSITIVE CONTROL REMAINED NEGATIVE. FOR THE CONTROL THEY USED BIOTESTCELL-P3. THE CUSTOMER DD NOT PROVIDE A PATIENT SAMPLE OR A PRODUCT SAMPLE OF THE ALLEGEDLY DEFECTIVE LOT OF SERACLONE ANTI-JKB FOR INVESTIGATIONAL TESTING. THEREFORE, OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLE ACCORDING TO THE INSTRUCTIONS FOR USE. TESTING SENSITIVITY AND TITERS: THE REPORTED LOT DEMONSTRATED HIGH SENSITIVITY, WITH TITERS EXCEEDING THE MINIMUM REQUIREMENT OF 1:8 ON TESTED JK(A+B+) DONOR SAMPLES. THE RESULTS WERE STRONG AND CONSISTENT WITH EXPECTATIONS. THE REFERENCE DONORS JK(A+B+) SHOWED REACTIVITY WITH TITERS OF 1:64 AND 1:32. TESTING SPECIFICITY: ALL JK(B+) RED CELLS (REFERENCE SAMPLES AND DONOR) REACTED STRONGLY POSITIVE. ALL JK(B-) CELLS REACTED CLEARLY NEGATI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22993044 ↗
  • MalfunctionDec 23, 2020False Negative Result; False Negative Result

    CUSTOMER REPORTED ABOUT FALSE NEGATIVE AND TOO WEAK POSITIVE REACTIONS WITH SERACLONE ANTI-JKB. THE CUSTOMER WAS RUNNING THE POSITIVE CONTROL OF THE SERACLONE ANTI-JKB FUSING MULTIPLE HETEROZYGOUS POSITIVE CELLS AND HE FOUND THE REACTIONS TO BE AMAZINGLY FRAGILE AND WEAK COMPARED TO PREVIOUS LOTS OF SERACLONE ANTI-JKB. THE CUSTOMER DECLARED THAT HE FOLLOWED THE PRODUCT IFU ACCORDING THE DIRECTIONS. THE CUSTOMER DID NEITHER PROVIDE THE COMPLAINT SAMPLE FOR INVESTIGATIONAL TESTING NOR THE RED BLOOD CELLS USED AS POSITIVE CONTROL. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLE OF SERACLONE ANTI-JK(B) (REF.#: 808184100, LOT #8937090-01, 2021-05-02) IN PARALLEL WITH LOT #8026020-01. BOTH LOTS WERE TESTED MANUALLY FOR POTENCY AND SPECIFICITY. FOR THE POTENCY AS WELL AS THE POSITIVE SPECIFICITY TESTING THE TESTS WERE INCUBATED FOR 15 MINUTES AT ROOM TEMPERATURE AND SUBSEQUENTLY CENTRIFUGED FOR 60 SECONDS AT 800 TO 1000 X G. THE MINIMUM TITER OF 8 WAS ALWAYS EXCEEDED. FOR THE TESTING OF POSITIVE SPECIFICITY JKA+B+ HETEROZYGOUS REAGENT RED BLOOD CELLS OF BIOTEST CELL AND PANOCELL (IMMUCOR) WERE USED. ALL POSITIVE REACTIONS WERE CORRECT AND STRONGLY POSITIVE. THE REAC

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11062525 ↗