IH-Card ABD(DVI-)-Conf 813181100Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Blood Grouping Reagent; Gel card for use with the IH-system; Card for ABD(DVI-) typing; 48 cards of 6 microtubes
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07611969952120), received between Sep 3, 2020 and Sep 29, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- False Positive Result3
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS OF CORD BLOOD SAMPLES WITH THE ANTI-B OF IH-CARD ABD(DVI-)-CONF ON THEIR IH-500 INSTRUMENT. THE CUSTOMER STATED THAT THEY RECEIVED CORRECT RESULTS IN THE TUBE TEST. THE CUSTOMER ANNOUNCED TO SEND IN THE COMPLAINT SAMPLE IH-CARD ABD(DVI-)-CONF FOR INVESTIGATIONAL TESTING. OUR QUALITY CONTROL LABORATORY HAS NOT YET STARTED TESTING BUT IS WAITING FOR THE COMPLAINT MATERIAL OF THE CUSTOMER. INSTRUMENT DATE OF THE AFFECTED IH-500 WERE REQUESTED AND THE INVESTIGATION OF THE RECEIVED DATABASES IS STILL ONGOING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17845525 ↗MalfunctionSep 29, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED FALSE POSITIVE REACTIONS OF CORD BLOOD SAMPLES WITH THE ANTI-B OF (B)(6) ON THEIR IH-500 INSTRUMENT. THE CUSTOMER STATED THAT THEY RECEIVED CORRECT RESULTS IN THE TUBE TEST. THE CUSTOMER ANNOUNCED TO SEND IN THE COMPLAINT SAMPLE (B)(6) FOR INVESTIGATIONAL TESTING. OUR QUALITY CONTROL LABORATORY HAS NOT YET STARTED TESTING BUT IS WAITING FOR THE COMPLAINT MATERIAL OF THE CUSTOMER. INSTRUMENT DATE OF THE AFFECTED IH-500 WERE REQUESTED AND THE INVESTIGATION OF THE RECEIVED DATABASES IS STILL ONGOING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17845516 ↗MalfunctionSep 3, 2020False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result
THE CUSTOMER REPORTED SEVERAL ISSUES WITH FALSE POSITIVE RESULTS INVOLVING IH-CARD ABD(DVI-)-CONF ON THE IH-1000. THE CUSTOMER STATED THAT THE SOFTWARE INTERPRETATION WAS 1+ POSITIVE, WHEN THE IMAGES WERE CLEARLY NEGATIVE. THE CUSTOMER DID NEITHER RETURN THE SUPPOSEDLY DEFECTIVE PRODUCT FOR INVESTIGATIONAL TESTING NOR THE PATIENT SAMPLE THAT HAD CAUSED THE FALSE POSITIVE TEST RESULT. THEREFORE OUR QUALITY CONTROL LABORATORY CHECKED THEIR RETENTION SAMPLE OF THE ALLEGEDLY DEFECTIVE PRODUCT LOT VISUALLY AND ALL ACCEPTANCE CRITERIA WERE MET. THE SPECIFICATIONS WERE: VISIBLE SUPERNATANT HOMOGENEOUS GEL INTACT SEALING NO SPLASHES IN THE REACTION CHAMBER FURTHERMORE, SEVERAL CARDS OF THE RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE PRODUCT IH-CARD ABD(DVI-)-CONF WERE TESTED WITH DONOR SAMPLES ON IH-1000. NO FALSE POSITIVE REACTIONS WERE OBSERVED. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. ACCORDING TO OUR SOP THE COMPLAINT WAS CLASSIFIED AS UNDETERMINED: THE COMPLAINT SAMPLE WAS NOT AVAILABLE, AND THE RETENTION SAMPLE REACTED AS EXPECTED. THE CUSTOMER PROVIDED IMAGES, BUT NOT FOR THI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10487283 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page