IH-Panel 11 Papain 814080100Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Reagent Red Blood Cells; 0.6 ± 0.1 %; For Use in identification of unexpected antibodies with IH-Cards; 11 x 4 mL
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07611969952304), received between Mar 22, 2019 and Mar 28, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2025Output Problem; Output Problem
THE CUSTOMER WANTED TO LOAD IH-PANEL 11 PAPAIN ON HIS IH-500 AND NOTICED IN CELLS 7 AND 2 SMALL CLUMPS OR THREADS. THEREFORE, NO TESTING COULD BE PERFORMED AND THE CUSTOMER SUSPECTED A POSSIBLE CONTAMINATION. THE CUSTOMER SENT IN THE ALLEGEDLY DEFECTIVE PRODUCT FOR INVESTIGATIONAL TESTING. ALL VIALS PROVIDED BY THE CUSTOMER WERE LEAKED OUT AS NO DROP PIPETTES WERE USED FOR SEALING OF THE OPENED VIALS. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLE OF THE AFFECTED LOT. THE CELLS WERE RESUSPEND AND IT WAS INVESTIGATED IF A PROPER RE-SUSPENSION IS GIVEN OR IF STRANDS OF CELLS ARE VISIBLE. AFTERWARDS THE FILLING VOLUME ON THE IH-500 WAS MEASURED. THE RETENTION SAMPLE WAS ALSO VISUALLY INSPECTED. WHEN CAREFULLY MIXED, NO SMALL CLUMPS OR THREADS WERE VISIBLE. THE SUSPENSIONS WERE WITHOUT ABNORMALITIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21717927 ↗MalfunctionMar 22, 2019False Negative Result
THE CUSTOMER REPORTED THAT THEY RAN THE IH-PANEL 11 PAPAIN IN PARALLEL TO THE UNTREATED PANEL (IH-PANEL 11) FOR THE SAME LOT OF REAGENTS. THE K+ CELL (#8) ON THE UNTREATED PANEL WAS TESTED AS 1+ AND THE SAME K+ CELL (CELL 8) ON THE IH-PANEL 11 PAPAIN YIELDED A NEGATIVE REACTION. THE RETENTION SAMPLE OF THE ALLEGEDLY DEFECTIVE LOF OF IH-PANEL 11 PAPAIN WAS TESTED WITHIN ITS SHELF LIFE IN OUR QUALITY CONTROL LABORATORY WITH AN ANTI-K AND THE RESULTS WERE AS EXPECTED CORRECT. THE PATIENT SAMPLE THAT HAD CAUSED A FALSE NEGATIVE REACTION WAS REQUESTED FOR INVESTIGATIONAL TESTING BUT HAS SO FAR NOT BEEN PROVIDED. TESTING IN OUR QUALITY CONTROL LABORATORY WILL BE RESUMED WITH TESTING OF THE PATIENT SAMPLE WITH THE CURRENT LOT OF IH-PANEL 11 PAPAIN AS SOON AS WE RECEIVE THE PATIENT SAMPLE. TESTING BY OUR QUALITY CONTROL LABORATORY CONFIRMED THAT THE ALLEGEDLY DEFECTIVE LOT OF IH-PANEL -I11 PAPAIN HAS DURING ITS SHELF LIFE FUNCTIONED CORRECTLY. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 8443421 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page