Biotestcell® A1 & B 816057100Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Reagent Red Blood Cells; Biotestcell® A1 & B; 3.0 - 3.4%; pooled cells for reverse grouping by the tube test; 2x10ml
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07611969952663), received between Apr 16, 2020 and May 21, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2024False Positive Result; False Positive Result
THE CUSTOMER REPORTED UNEXPECTED POSITIVE REACTIONS OF PATIENT SAMPLES WITH BIOTESTCELL A1&B IN THE REVERSE TYPING. THE CUSTOMER STATED THAT SHE A PATIENT SAMPLE WITH AN ANTI-M CAUSED INTERFERENCES WITH BIOTESTCELL A1&B. FURTHERMORE, THE CUSTOMER STATED THAT THE REVERSE TYPING REAGENT RED BLOOD CELLS OF ORTHO WOULD BE M ANTIGEN NEGATIVE. THE DUSTOMER DID NOT PROVIDE A DATE OF EVENT. THE CUSTOMER DID NEITHER RETURN A PRODUCT SAMPLE OF BIOTESTCELL A1&B, LOT 9402011-00 FOR INVESTIGATIONAL TESTING NOR THE PATIENT SAMPLE THAT SHOWED INTERFERENCES. THEREFORE, OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLES OF THE CURRENT LOTS OF BIOTESTCELL A1&B: LOT 9402011-00 (LOT THE CUSTOMER FILED HER COMPLAINT FOR), LOT 9406011-00 AND LOT 9410011-00 WITH DIFFERENT DONOR SAMPLES ACCORDING TO THE INSTRUCTION FOR USE (IFU). ALL ACCEPTANCE CRITERIA WERE MET. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTION. WITHIN THE PRODUCTION PROCESS BIOTESTCELL A1&B WAS TESTED FOR THE ABO BLOODGROUPS, THE RH PHENOTYPES AND THE ANTIGENS KELL (KEL1) AND CELLANO (KEL2) ACCORDING TO THE REQUIREMENTS. THE ACCEPTANCE CRITERIA WERE MET. TESTING FOR FURTHER ANTIGENS WAS NOT MANDATORY. BASED ON THE INVESTIGATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19357772 ↗MalfunctionApr 16, 2020False Positive Result
THE CUSTOMER REPORTED ISSUES WITH BIOTESTCELL A1&B. THE CUSTOMER STATED THAT THERE ARE PATIENTS THAT GENERATED WHAT SHE DESCRIBED AS ROULEAUX FORMATION IN THE REVERSE TYPING. THE CUSTOMER STATED THAT SHE SAW THE EXTRA REACTION (FALSE POSITIVE) REACTION WITH BIOTESTCELL A1. FURTHERMORE SHE REPORTED THAT THEY USED SUCCESSFULLY A SALINE REPLACEMENT ACCORDING TO THE AABB TECHNICAL MANUAL TO RESOLVE UNEXPECTED POSITIVE REACTIONS. THE EXACT DATE OF EVENT IS NOT KNOWN. THE CUSTOMER HAS NOT RETURNED THE SUPPOSEDLY DEFECTIVE PRODUCT BIOTESTCELL A1&B FOR INVESTIGATIONAL TESTING BUT WILL PROVIDE TWO PATIENT SAMPLES THAT HAD CAUSED FALSE POSITIVE TEST RESULTS. IN THE MEAN TIME OUR QUALITY CONTROL LABORATORY TESTED THEIR RETENTION SAMPLE WITH DIFFERENT DONOR SAMPLES IN THE TUBE TECHNIQUE. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTION RESPECTIVELY ROULEAUX FORMATION. ADDITIONALLY THE BLOOD GROUP A RESPECTIVELY B OF THE RETENTION SAMPLE WAS CONFIRMED WITH THE BLOOD GROUPING REAGENTS SERACLONE ANTI-A, SERACLONE ANTI-B, SERACLONE ANTI-AB, SERACLONE ANTI-A1 AND SERACLONE ANTI-H. AGAIN, ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. A REVIEW OF TH
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9966960 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page