IH-Card Anti-D (DVI+) 813251100Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Blood Grouping Reagent; Gel card for use with the IH-system; Card for single D (RH1) (DVI+) antigen typing; 48 cards of 6 microtubes
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07611969962402), received between Aug 1, 2022 and Sep 5, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- False Positive Result3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2022False Positive Result; False Positive Result
THE CUSTOMER REPORTED WEAK POSITIVE RESULTS WITH THE ANTI-D INSTEAD OF CLEARLY NEGATIVE RESULTS ON IH-1000 WHEN USING THE IH-CARD ANTI-D (DVI+) . ORIGINALLY, THE CUSTOMER STATED THAT HE OBSERVED THIS ISSUE WITH HEMOBIOSCIENCE CONTROL QC3 WHICH IS SUPPOSED TO BE RHD NEGATIVE. THE CUSTOMER PROVIDED A PRODUCT SAMPLE OF THE AFFECTED IH-CARD ANTI-D (DVI+) FOR INVESTIGATIONAL TESTING . OUR QUALITY CONTROL LABORATORY IS CURRENTLY TESTING THE COMPLAINT SAMPLE IN PARALLEL WITH THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT ON IH-1000. THIS TESTING IS ONGOING. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. ALSO STILL ONGOING IS AN ANALYSIS OF THE TRACE FILES OF THE AFFECTED INSTURMENT IH-1000.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15354461 ↗MalfunctionSep 5, 2022False Positive Result; False Positive Result
THE CUSTOMER REPORTED WEAK POSITIVE RESULTS WITH THE ANTI-D INSTEAD OF CLEARLY NEGATIVE RESULTS ON IH-1000 WHEN USING THE IH-CARD ANTI-D (DVI+) . ORIGINALLY, THE CUSTOMER STATED THAT HE OBSERVED THIS ISSUE WITH HEMOBIOSCIENCE CONTROL QC3 WHICH IS SUPPOSED TO BE RHD NEGATIVE. THE CUSTOMER PROVIDED A PRODUCT SAMPLE OF THE AFFECTED IH-CARD ANTI-D (DVI+) FOR INVESTIGATIONAL TESTING . OUR QUALITY CONTROL LABORATORY IS CURRENTLY TESTING THE COMPLAINT SAMPLE IN PARALLEL WITH THEIR RETENTION SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT ON IH-1000. THIS TESTING IS ONGOING. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVELY AFFECTED THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT. ALSO STILL ONGOING IS AN ANALYSIS OF THE TRACE FILES OF THE AFFECTED INSTRUMENT IH-1000.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15354460 ↗MalfunctionAug 1, 2022False Positive Result; False Positive Result
THE CUSTOMER REPORTED DRIED GEL COLUMNS IN POSITION 6 OF THE IH-CARD ANTI-D (DVI+) AND STATED THAT RESULTS OF KNOWN RHD NEGATIVE SAMPLES WERE WEAK POSITIVE ON THE IH-1000. THE CUSTOMER PROVIDED AN IMAGE OF AN UNPROCESSED CARD WHICH SHOWED A LOWER FILLING VOLUME IN COLUMN 6. ADDITIONALLY, THE CUSTOMER PROVIDED AN IMAGE OF THE DAILY JOURNAL THAT SHOWED A 1+ POSITIVE REACTION IN THE ANTI-D. THE IMAGE SHOWED A CLEARLY WEAK POSITIVE REACTION, BUT SIGNS OF DRYING WERE NOT VISIBLE IN THE AFFECTED GEL COLUMN. THE CUSTOMER RETURNED 87 IH-CARDS ANTI-D(DVI+) FOR INVESTIGATIONAL TESTING. OUR QUALITY CONTROL LABORATORY VISUALLY CHECKED THEM FOR INTACT SEALING, HOMOGENEOUS GEL, VISIBLE SUPERNATANT, AND ABSENCE OF SPLASHES IN THE REACTION CHAMBER. ALL ACCEPTANCE CRITERIA WERE MET. THE SUPERNATANT WAS REDUCED, BUT STILL VISIBLE. IT WAS KNOWN AND ACCEPTED THAT THE SUPERNATANT TOWARDS THE END OF THE SHELF LIFE WILL BE VERY, VERY SMALL. WITHIN OUR STABILITY STUDY WE COULD CONFIRM THAT THESE CARDS WORKED PROPERLY AND FULFILLED THE ACCEPTANCE CRITERIA. ADDITIONALLY, CARDS RETURNED BY THE CUSTOMER WERE TESTED WITH DONOR SAMPLES ON THE IH1000. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. A REVIEW OF
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15137148 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page