IH-Card ABD(DVI-)-Conf 813182100Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Blood Grouping Reagent; Gel card for use with the IH-system; Card for ABD(DVI-) typing; 288 cards of 6 microtubes
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07611969964604), received between Nov 20, 2019 and Nov 20, 2019 — classified as 5 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- False Negative Result5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2019False Negative Result
THE CUSTOMER MADE A GENERAL COMPLAINT DURING A MEETING AT THE AABB CONVENTION IN 2019 AND REPORTED ABO DISCREPANCIES ON IH-1000. THE CUSTOMER STATED THAT ONE OF THE WELLS OF THE BACK TYPE HAD NO REACTION WHEN A POSITIVE REACTION WAS EXPECTED. FURTHERMORE, THE CUSTOMER STATED THAT THE REACTIONS WERE CORRECTLY POSITIVE IN THE TUBE TEST. THE CUSTOMER HAS OBSERVED THIS ISSUE WITH SINGLE PATIENT SAMPLES SINCE JULY 2019 BUT VOICED HIS COMPLAINED ONLY IN OCTOBER 2019 . THE CUSTOMER PROVIDED IH-1000 IMAGES OF SIX SPECIMENS WHICH CONFIRMED THE FALSE NEGATIVE REACTIONS IN THE REVERSE TYPING. THESE SIX SAMPLES WERE TESTED ON JULY13 (ONE SPECIMEN), JULY 15 (ONE SPECIMEN), JULY 18 (ONE SPECIMEN), JULY 22 (TWO SPECIMEN) AND JULY 24 (ONE SPECIMEN), 2019. DUE TO THE DISCREPANCY BETWEEN FORWARD AND REVERSE TYPING THE IH-1000 STATED "ABO NOT INTERPRETABLE" AND NO INCORRECT RESULTS WERE RELEASED. BECAUSE OF THE TIME LAPSE BETWEEN OBSERVING THESE FALSE NEGATIVE REAGCTIONS AND FILING HIS COMPLAINT, THE CUSTOMER WAS NOT ABLE TO RETURN THE SUPPOSEDLY DEFECTIVE PRODUCT, THE AFFECTED PATIENT DONOR SAMPLES THAT HAD REACTED FALSE NEGATIVE IN REVERSE TYPING AS WELL AS THE NEEDED TRACE FILES FOR FURTHER INVEST
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9346457 ↗MalfunctionNov 20, 2019False Negative Result
THE CUSTOMER MADE A GENERAL COMPLAINT DURING A MEETING AT THE RECENT (B)(6) CONVENTION. THE CUSTOMER REPORTED ABO DISCREPANCIES ON IH-1000. HE STATED THAT ONE OF THE WELLS OF THE BACK TYPE HAD NO REACTION WHEN A POSITIVE REACTION WAS EXPECTED. FURTHERMORE HE STATED THAT THE REACTIONS WERE CORRECTLY POSITIVE IN THE TUBE TEST. THE CUSTOMER HAS BEEN OBSERVING THIS ISSUE ON SINGLE PATIENT SAMPLES SINCE (B)(6), BUT JUST MENTIONED THIS ISSUE DURING THE (B)(6) ON (B)(6) 2019. THE CUSTOMER PROVIDED IH-1000 IMAGES OF SIX SPECIMENS WHICH HAD CAUSED ALLEGEDLY FALSE NEGATIVE REACTIONS IN THE REVERSE TYPING. THESE SIX SAMPLES WERE TESTED ON (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (TWO SPECIMEN) AND (B)(6) (ONE SPECIMEN), (B)(6). WE ARE STILL TRYING TO GET ALL MISSING INFORMATION LIKE E.G. LOT NUMBER AND CONCOMITANT PRODUCTS FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9346456 ↗MalfunctionNov 20, 2019False Negative Result
THE CUSTOMER MADE A GENERAL COMPLAINT DURING A MEETING AT THE AABB CONVENTION IN 2019 AND REPORTED ABO DISCREPANCIES ON IH-1000. THE CUSTOMER STATED THAT ONE OF THE WELLS OF THE BACK TYPE HAD NO REACTION WHEN A POSITIVE REACTION WAS EXPECTED. FURTHERMORE, THE CUSTOMER STATED THAT THE REACTIONS WERE CORRECTLY POSITIVE IN THE TUBE TEST. THE CUSTOMER HAS OBSERVED THIS ISSUE WITH SINGLE PATIENT SAMPLES SINCE (B)(6) 2019 BUT VOICED HIS COMPLAINED ONLY IN (B)(6) 2019 . THE CUSTOMER PROVIDED IH-1000 IMAGES OF SIX SPECIMENS WHICH CONFIRMED THE FALSE NEGATIVE REACTIONS IN THE REVERSE TYPING. THESE SIX SAMPLES WERE TESTED ON (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (TWO SPECIMEN) AND (B)(6) (ONE SPECIMEN), 2019. DUE TO THE DISCREPANCY BETWEEN FORWARD AND REVERSE TYPING THE IH-1000 STATED "ABO NOT INTERPRETABLE" AND NO INCORRECT RESULTS WERE RELEASED. BECAUSE OF THE TIME LAPSE BETWEEN OBSERVING THESE FALSE NEGATIVE REAGCTIONS AND FILING HIS COMPLAINT, THE CUSTOMER WAS NOT ABLE TO RETURN THE SUPPOSEDLY DEFECTIVE PRODUCT, THE AFFECTED PATIENT DONOR SAMPLES THAT HAD REACTED FALSE NEGATIVE IN REVERSE TYPING AS WELL AS THE NEEDED TRACE FILES FOR FURTHER INVESTIGA
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9346455 ↗MalfunctionNov 20, 2019False Negative Result
THE CUSTOMER MADE A GENERAL COMPLAINT DURING A MEETING AT THE AABB CONVENTION IN 2019 AND REPORTED ABO DISCREPANCIES ON IH-1000. THE CUSTOMER STATED THAT ONE OF THE WELLS OF THE BACK TYPE HAD NO REACTION WHEN A POSITIVE REACTION WAS EXPECTED. FURTHERMORE, THE CUSTOMER STATED THAT THE REACTIONS WERE CORRECTLY POSITIVE IN THE TUBE TEST. THE CUSTOMER HAS OBSERVED THIS ISSUE WITH SINGLE PATIENT SAMPLES SINCE (B)(6) 2019 BUT VOICED HIS COMPLAINED ONLY IN OCTOBER 2019 . THE CUSTOMER PROVIDED IH-1000 IMAGES OF SIX SPECIMENS WHICH CONFIRMED THE FALSE NEGATIVE REACTIONS IN THE REVERSE TYPING. THESE SIX SAMPLES WERE TESTED ON (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (TWO SPECIMEN) AND (B)(6) (ONE SPECIMEN), 2019. DUE TO THE DISCREPANCY BETWEEN FORWARD AND REVERSE TYPING THE IH-1000 STATED "ABO NOT INTERPRETABLE" AND NO INCORRECT RESULTS WERE RELEASED. BECAUSE OF THE TIME LAPSE BETWEEN OBSERVING THESE FALSE NEGATIVE REACTIONS AND FILING HIS COMPLAINT, THE CUSTOMER WAS NOT ABLE TO RETURN THE SUPPOSEDLY DEFECTIVE PRODUCT, THE AFFECTED PATIENT DONOR SAMPLES THAT HAD REACTED FALSE NEGATIVE IN REVERSE TYPING AS WELL AS THE NEEDED TRACE FILES FOR FURTHER INVESTIGA
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9346454 ↗MalfunctionNov 20, 2019False Negative Result
THE CUSTOMER MADE A GENERAL COMPLAINT DURING A MEETING AT THE RECENT AABB CONVENTION. THE CUSTOMER REPORTED ABO DISCREPANCIES ON IH-1000. HE STATED THAT ONE OF THE WELLS OF THE BACK TYPE HAD NO REACTION WHEN A POSITIVE REACTION WAS EXPECTED. FURTHERMORE HE STATED THAT THE REACTIONS WERE CORRECTLY POSITIVE IN THE TUBE TEST. THE CUSTOMER HAS BEEN OBSERVING THIS ISSUE ON SINGLE PATIENT SAMPLES SINCE (B)(6), BUT JUST MENTIONED THIS ISSUE DURING THE AABB ON (B)(6) 2019. THE CUSTOMER PROVIDED IH-1000 IMAGES OF SIX SPECIMENS WHICH HAD CAUSED ALLEGEDLY FALSE NEGATIVE REACTIONS IN THE REVERSE TYPING. THESE SIX SAMPLES WERE TESTED ON (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (ONE SPECIMEN), (B)(6) (TWO SPECIMEN) AND (B)(6) (ONE SPECIMEN), 2019. WE ARE STILL TRYING TO GET ALL MISSING INFORMATION LIKE E.G. LOT NUMBER AND CONCOMITANT PRODUCTS FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9346453 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page