Bromelin for Erytype 806210100 (UDI 07611969965755)Recalls & reported problems
Bio-Rad Medical Diagnostics GmbH · Red Blood Cell Diluent for TANGO instruments, 10x10 mL · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07611969965755), received between Jul 27, 2017 and Aug 11, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- False Positive Result2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2017False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULTS FOR ANTI-D ON TWO DIFFERENT TANGO OPTIMO ON SITE WHEN USING BROMELIN FOR ERYTYPE. THE CUSTOMER SUSPECTED THAT THE BROMELIN FOR ERYTYPE WOULD NOT WORK FOR FULL 24 HOURS AFTER PREPARING THE WORKING SOLUTION. THE CUSTOMER OBSERVED "WEIRD LOOKING RESULTS" IN THE D- WELLS WHEN USING SOLUTION THAT WAS PREPARED APPROXIMATELY 22 HOURS AGO. THE CUSTOMER DID NOT SEND IN THE ALLEGEDLY DEFECTIVE FOR INVESTIGATIONAL TESTING. THEREFORE OUR QUALITY CONTROL LABORATORY TESTED THEIR RETAINED SAMPLE OF THE SUPPOSEDLY DEFECTIVE LOT BROMELIN FOR ERYTYPE AND ALSO A REFERENCE LOT. BOTH REAGENTS WERE TESTED WHEN FRESHLY PREPARED AND A SECOND TIME AFTER 22 HOURS "ON BOARD". ERYTYPE S ABD+REV.A1B #8646110 WAS USED FOR THE TESTING WHICH IS THE SAME LOT THAT WAS USED BY THE CUSTOMER. ALL POSITIVE AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTION. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVE INFLUENCES ON THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT BROMELIN FOR ERYTYPE. BOTH AFFECTED TANGO OPTIMO WERE CHECKED BY OUR FIELD SERVICE ENGINEER. NO INDICATION FOR AN INSTRUMENT MALFUNCTION
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6787396 ↗MalfunctionJul 27, 2017False Positive Result
THE CUSTOMER REPORTED A DISCREPANCY WHEN TYPING ON TWO TANGO OPTIMO INSTRUMENTS. RH NEGATIVE SAMPLES WERE BEING READ AS POSITIVE. THE CUSTOMER SUSPECTED THE BROMELIN FOR ERYTYPE TO BE THE CAUSE FOR THE FALSE POSITIVE RESULTS. THE CUSTOMER SENT BACK THE ALLEGEDLY DEFECTIVE PRODUCT BROMELIN FOR ERYTYPE FOR INVESTIGATIONAL TESTING BUT AT THE TIME THE ALLEGEDLY DEFECTIVE PRODUCT ARRIVED AT BMD IT WAS ALREADY EXPIRED, WAS CONTAMINATED AND EVEN MOULDY. THEREFORE OUR QUALITY CONTROL LABORATOY TESTED THEIR RETAINED SAMPLE BROMELIN FOR ERYTYPE, #8701110-00. THE ERYTYPE S ABD+REV. A1B #8646110, THAT WAS USED BY THE CUSTOMER, WAS ALSO USED IN THIS TEST. 12 SAMPLES WERE TESTED AND ALL REACTIONS WERE AS EXPECTED. WE DID NOT OBSERVE ANY FALSE POSITIVE REACTION. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE NEGATIVE INFLUENCES ON THE QUALITY OF THE ALLEGEDLY DEFECTIVE LOT OF BROMELIN FOR ERYTYPE. BOTH AFFECTED TANGO OPTIMO WERE INSPECTED BY OUR FIELD SERVICE ENGINEER WITH NO INDICATION FOR A MALFUNCTION. BOTH INSTRUMENTS ARE WORKING WITHIN SPECIFICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6747023 ↗
Other System, Test, Automated Blood Grouping And Antibody devices
- Galileo Echo 0087000Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody242 reports
- ORTHO 6904577Ortho-Clinical Diagnostics, Inc. · SYSTEM, TEST, AUTOMATED BLOOD GROUPING AND ANTIBODY117 reports
- Galileo Neo 0064599Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody65 reports
- TANGO infinity 850000010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody1 death55 reports
- Echo Lumena 0086998Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody49 reports
- IH-1000 001100Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody44 reports
- NEO Iris 0064597Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody28 reports
- TANGO optimo 848900010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody24 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.