SL-PLUS 75004613Recalls & reported problems
Smith & Nephew, Inc. · PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07611996085471), received between Jan 9, 2020 and Apr 17, 2025 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A THR SURGERY, ONE (1) PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM BROKE DURING THE FEMORAL BROACHING. IT IS UNKNOWN IF ANY PIECE FELL INSIDE THE INCISION. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21862220 ↗InjuryJan 18, 2022Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE END OF A PLUS SL MIA DBL OFFSET ADAPT. RT 60/25MM, THAT CONNECTS TO THE STRIKE PAD, BROKE OFF. IT IS UNKNOWN IF ANY PIECES FELL INSIDE OF THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13280623 ↗InjuryDec 21, 2021Break; Break
IT WAS REPORTED THAT, DURING TOTAL HIP REPLACEMENT, A PLUS SL-PLUS MIA DBL OFFSET ADAPT. RIGHT 60/25MM BROKE ON IMPACTION, INSIDE THE PATIENT. ALL PARTS WERE RETRIEVED BY HAND. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13054835 ↗InjuryDec 10, 2021Break; Break
IT WAS REPORTED THAT, DURING TOTAL HIP REPLACEMENT, THE END OF THE HANDLE OF A SL- PLUS MIA DBL OFFSET ADAPT. RT 60/25MM BROKE OFF AT THE STRIKE PLATE CONNECTION, INSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12976215 ↗MalfunctionDec 10, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE PLUS SL MIA DBL OFFSET ADAPT. RT 60/25 MM BROKE WHILE IMPACTING. SURGERY WAS RESUMED, WITH A BACK-UP DEVICE. IT IS UNKNOWN IF THERE WAS A DELAY AS CONSEQUENCE OF THIS ALLEGATION. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12976135 ↗MalfunctionMay 13, 2021Break; Break
IT WAS REPORTED THAT, DURING THR SURGERY, THE CONNECTION POST FOR STRIKE PLATE OF A POLARSTEM BROACH HANDLE SNAPPED OFF OUTSIDE THE PATIENT. NO SIGNIFICANT SURGICAL DELAY WAS REPORTED AS A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NEITHER PATIENT INJURY NOR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821071 ↗MalfunctionFeb 21, 2020Break; Break
IT WAS REPORTED THAT THE BROACH HANDLE BROKE UPON IMPACTION. A PIECE FELL INTO THE PATIENT'S WOUND AND IT WAS SUCCESSFULLY RECOVERED. S+N BACKUP DEVICE WAS AVAILABLE AND NO DELAY WAS RECORDED.
Patient problems: No Information
MAUDE report 9738322 ↗MalfunctionJan 9, 2020Break; Break
IT WAS REPORTED THAT THE CONNECTOR OF OFFSET BROACH HANDLE BROKE UPON IMPACTION. ALL PARTS RECOVERED WITHOUT ISSUE. DELAY NO GREATER THAN THIRTY MINUTES WAS REPORTED. BACKUP DEVICE WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9568469 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page