POLARSTEM 75004678Recalls & reported problems
Smith & Nephew, Inc. · POLARSTEM EXTRACTOR BLOCK · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07611996109399), received between Jan 14, 2021 and Jun 12, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture2
- Break1
- Material Deformation1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT, DURING TOTAL HIP REPLACEMENT, A POLARSTEM EXTRACTOR BLOCK BROKE INSIDE THE PATIENT UPON EXTRACTION. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A COMPETITOR'S DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14672467 ↗MalfunctionMay 7, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING FIELD INSPECTION THE CS/CSL/GERADSCHAFT-PLUS EXTRACT.SCREW M6 AND THE POLARSTEM EXTRACTOR BLOCK WERE STRIPPED AND DID NOT ENGAGE PROPERLY. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11793795 ↗InjuryJan 14, 2021Fracture; Material Deformation; Fracture; Material Deformation
IT WAS REPORTED THAT THE POLARSTEM EXTRACTOR BLOCK BROKE WHILE TRYING TO SLAP OUT AN IMPLANTED POLAR STEM. THE PROCEDURE WAS COMPLETED USING A S+N BACKUP DEVICE. NO INJURY TO THE PATIENT WAS REPORTED. A SURGICAL DELAY OF 30 MIN OR LESS WAS REPORTED. THE NEWLY IMPLANTED STEM HAD TO BE EXTRACTED BECAUSE THE SURGEON WAS NOT ABLE TO REDUCE THE HIP AFTER IMPLANTING. AS A RESULT, THE STEM WAS DOWNSIZED AND SAT LOWER IN THE CANAL, CREATING ROOM FOR THE HIP TO BE REDUCED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11169250 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page