POLARCUP 75100440Recalls & reported problems
Smith & Nephew, Inc. · POLARCUP SHELL TI-PLASMA/HA 51 NON-CEMENTED · Class 2 · Implant
All models of POLARCUP (61)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07611996118315), received between Mar 10, 2021 and May 7, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K122244, K211176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2026Mechanical Jam
IT WAS REPORTED THAT, DURING AN ACETABULAR REVISION SURGERY (UNKNOWN IF FROM S+N IMPLANTS), ON ATTEMPTING TO UNSCREW ONE (1) POLARCUP ANCHORING PEG 2 EACH OF THE PEG HOLE COVERS FROM ONE (1) POLARCUP SHELL TI-PLASMA/HA 51 NON-CEM, THE THREADS CROSS THREADED AND COULD NOT BE REMOVED. THE SURGEON ATTEMPTED TO SEAT A PEG IN THE SECOND, AVAILABLE PEG HOLE AFTER DRILLING THE HOLE AS PER THE TECHNIQUE, BUT THE PEG FAILED TO SEAT FLUSH WITH THE CUP AND HAD TO BE REMOVED. THE PEG HOLE WAS LEFT OPEN. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25112276 ↗MalfunctionMay 12, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING B/L THR, ONE PLUG COVER OF THE POLARCUP SHELL TI-PLASMA/HA 51 NON-CEM COULD NOT BE REMOVED AND THE POLARCUP SHELL TI-PLASMA/HA 51 NON-CEM SUFFERED A DEFORMATION WHILE TRYING TO UNPLUGED THE COVER. NO SIGNIFICANT DELAYS WERE REPORTED AND THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. NO PATIENT WAS HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11814116 ↗MalfunctionMar 10, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT WHEN DOCTOR OPENED POLARCUP SHELL, HE OPENED THE SCREWS TO GET PEGS. NUTS ON THE POLARCUP SHELL CANNOT BE OPENED DESPITE ATTEMPT OF UNSCREWING AND SCREWING FOR OPENING. THE PROCEDURE WAS COMPLETED USING A SN BACKUP DEVICE. NO SURGICAL DELAY OR INJURY TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11450006 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122244, K211176. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page